background image

42

Starting up

Use only high quality leak-proof batter-
ies of the correct specifications (see
“Technical data” P. 59).

 

Never

mix old and new batteries or

different makes.

 

If the batteries are not inserted cor-

rectly, the monitor will not work and this
may cause the batteries to heat up and
leak, damaging the monitor.

Inserting the batteries

The battery compartment is located on
the underside of the monitor. Insert bat-
teries in the battery compartment as
shown in illustration 1.

Remove the batteries if the monitor is
not to be used for any length of time.

+

+

+

+

Illustration 1

The device must be installed and put into
operation in accordance with the instruc-
tions in this manual.

Содержание CA02

Страница 1: ...medicus Gebrauchs anweisung User Instructions Mode d emploi Manuale di istruzioni Manual del Usuario 쮕 2 33 64 95 126 ...

Страница 2: ... values 40 Systolic and diastolic blood pressure value 40 WHO blood pressure cut off values 40 Starting up 42 Inserting batteries 42 Battery status 43 Care for the environment 43 General instructions for self measurement 44 Preparing to measure your blood pressure 46 Note on cuff sizes 46 Attaching the cuff 48 ...

Страница 3: ...rement 52 Measurement display 53 Mains operation 55 Error messages 56 Cleaning and Disinfection 57 Warranty 57 Customer service Disposal 58 Technical data 59 Calibration checks Testing instructions for trained specialist staff only 62 34 Table of contents ...

Страница 4: ...35 Mains connector pressure selector switch START STOP button LCD display Cuff connector Product features Battery status display see page 43 Measurement in progress see page 52 ...

Страница 5: ... une preuve d achat facture Este certificado de garantía es válido sólo en combinación con una prueba de compra factura Questo certificato di garanzia è valido solo in combinazione con una prova di acquisto fattura GARANTIE URKUNDE CERTIFICATE OF WARRANTY CERTIFICAT DE GARANTIE CERTIFICADO DE GARANTIA CERTIFICATO DI GARANZIA medicus Gebrauchs anweisung User Instructions Mode d emploi Manuale di is...

Страница 6: ... by pressing the START STOP button P 51 Measured systolic diastolic and pulse values are displayed after the measure ment is completed P 53 This quick guide does not replace the detailed information on the use and safety of the blood pressure monitor fea tured in the user instructions Please be sure therefore to read the user instructions ...

Страница 7: ...i able partner for monitoring your blood pressure values Please read these user instructions before using the instrument for the first time as correct blood pressure readings can only be obtained if the instrument is operated correctly In these user instructions means an action by the user To get help concerning usage or mainten ance please contact your dealer or the manufacturer contact details c...

Страница 8: ...icus blood pressure monitor is suitable for patients of all ages with an upper arm circumference of be tween 22 and 48 cm The monitor is not suitable for neonatals Wireless communication devices such as home networking devices mobile phones cordless phones and their base stations walkie talkies can affect this blood pres sure monitor Therefore a minimum distance of 11 feet should be kept from such...

Страница 9: ...eartmuscleex pands and fills with blood again Blood pressure measurement values are expressed in mmHg mm of mercury systolic diastolic High blood pressure hypertension over 140 mmHg over 90 mmHg Borderline normal blood pressure 130 to 139 mmHg 85 to 89 mmHg Normal blood pressure 120 to 129 mmHg 80 to 84 mmHg Optimum blood pressure up to 119 mmHg up to 79 mmHg The World Health Organization WHO has ...

Страница 10: ...e of the val ues systole diastole consistently exceeds the cut off point for high blood pressure Self measurement of blood pressure often produces somewhat lower values at home than at the doctor s The German Hypertension League has therefore defi ned lower cut off values for home measu rements Home measurement 135 85 mmHg Measurement in the doctor s surgery 140 90 mmHg ...

Страница 11: ...k and this may cause the batteries to heat up and leak damaging the monitor Inserting the batteries The battery compartment is located on the underside of the monitor Insert bat teries in the battery compartment as shown in illustration 1 Remove the batteries if the monitor is not to be used for any length of time Illustration 1 The device must be installed and put into operation in accordance wit...

Страница 12: ...shing No further measurements possible Replace battery Care for the environment Do not dispose of used batteries and charger units in domestic waste You can take these to collection sites for used batteries or special waste Contact your local authority for informa tion ...

Страница 13: ... prescribed by your doctor of your own accord 3 Irregular heartbeats can affect the accu racy of the monitor or result in incorrect measurements 4 Incorrect measurements can also occur in patients with a weak pulse possibly in patients wearing a pacemaker for ex ample The blood pressure monitor it self has no effect on the pacemaker 5 If you are pregnant consult your doctor before using the monito...

Страница 14: ...45 General instructions for self measurement To do this measure the blood pressure first in both arms and thereafter always from the arm with the higher blood pressure ...

Страница 15: ...er arm cir cumference printed on them Cuff supplied as standard Type Arm Order number circumference CA01 22 32 cm 143 4 760 CA02 32 48 cm 143 4 757 Preparing to measure your blood pressure Attach the cuff to the monitor by pressing the plug click into the socket on the left side of the unit See illustration 2 Illustration 2 ...

Страница 16: ...re which should be at least 30 to 40 mmHg higher than the systolic upper blood pressure reading If you are not sure of this value set the selec tor switch to 150 mmHg for the first measurement If this pressure is insufficient the unit will switch over to the next higher pres sure setting automatically Illustration 3 ...

Страница 17: ...aking a measurement Do not move during the measure ment Attaching the cuff The measurement should be taken from the bare upper arm In the case of tightly fitting outer clothing ensure that the blood supply to the arm is not cut off when the sleeve is rolled up if necessary remove the article of clothing before taking the measurement Open out the cuff into the shape of a ring and slide it over the ...

Страница 18: ...ated side of the body During the measurement medical ap paratus that are being used on the same arm at the same time may malfunction The cuff must not be too tight There should be enough space to fit roughly two fingers between arm and cuff The metal ring should never lie over the artery see illustration 4 as this may produce a false reading Grasp the end of the cuff that feeds though the metal ri...

Страница 19: ...low values too high values Illustration 6 Check that the air tube is not kinked during the measurement This could cause congestion of the blood flow and sub sequent injury The blood pressure measurement should not prevent the flow of blood for an unnecessary length of time 2 minutes If the monitor fails to function correctly re move the cuff from the arm Excessive frequent measurements can cause i...

Страница 20: ...ement with the START STOP button Now hold the arm absolutely still and do not talk All the items on the LCD display appear briefly to indicate that the monitor is now ready for use see illustration 6 Illustration 6 Systole mmHg Diastole mmHg Pulse min ...

Страница 21: ...ic signal sounds once the prese lected pressure has been reached The pump switches off and the air in the cuff escapes automatically Interrupting a measurement A measurement can be interrupted at any stage by pressing the start button the cuff deflates automatically During the measurement phase the fall ing pressure and the symbol are indi cated on the display Illustration 7 ...

Страница 22: ...ters the first beats the symbol blinks in time with the pulse rhythm and an acoustic signal can be heard The end of the measurement is indicated by a long beeping sound At the same time the built in solenoid valve opens automa tically to rapidly release the remaining pressure in the cuff Systole mmHg Diastole mmHg Pulse min Illustration 8 ...

Страница 23: ...ress disease etc If the measurement result is obviously wrong repeat the measurement The unit switches itself to time display mode automatically after about 1 min To repeat a measurement press the START STOP button again briefly Allow at least two minutes be tween measurements If no further measurement is required remove the cuff from the arm Enter each individual value in your record card 2 min ...

Страница 24: ...oso power supply unit order number 410 7 150 This power supply unit produces a rectified output of the correct polarity Other com mercially available power supply units can cause damage to the electronic compo nents which will invalidate the warranty There is also a risk of fire with these other power supply units ...

Страница 25: ...ading Meaning of error messages Fehlerbehebung Inflation pressure inadequate after Select the next higher pressure two attempts to inflate setting No pulses were detected that Check the positioning of the cuff could be evaluated Leak in the air system Check air connector on cuff Cuff possibly defective Fehlerursache für Err Anzeige ...

Страница 26: ...we recommend the disinfectant mikrozid sensitive liquid Schülke Mayr To disinfect the velcro of the cuff we re commend spray disinfection In particular if the device is used on multi ple users is to pay attention to regular cleaning and disinfections of the cuff Warranty For details of the warranty conditions please consult the warranty certificate supplied A claim under the warranty will only be ...

Страница 27: ...ted operational lifetime of the equip ment 10 years Expected operational lifetime of the cuff 10 000 measurement cycles Customer service Customer service Warranty and repair work must be under taken by trained and authorised personnel Do not modify this equipment without aut horization of the manufacturer Send the monitor carefully packaged and with sufficient postage to your authorised dealer or ...

Страница 28: ...10 C to 40 C Relative humidity 15 to 85 Storage conditions environmental temperature 10 C to 60 C Relative humidity 15 to 85 Power supply DC 6V 4 x 1 5V Mignon IEC LR 6 alkaline manganese batteries Alternative special option DC 6V power supply unit Order No 410 7 150 Typical battery life 1 000 measurement cycles depending on inflation pressure and frequency of use Technical data ...

Страница 29: ...mm x 70 mm x 110 mm Classification Protection class II Type BF Protection against solid IP20 objects Clinical test Accuracy complies with the requirements DIN 58130 of EN 1060 Part 3 Maximum deviation of cuff 3 mmHg pressure measurement Maximum deviation of 5 pulse rate display 60 Technical data ...

Страница 30: ...61 Applicable standards EN 1060 Part 1 Non invasive sphygmomanometers General requirements and Part 3 Supplementary requirements for electro mechanical blood pressure measuring systems Technical data ...

Страница 31: ...n of pressure display Calibration checks Testing instructions for trained specialist staff only Note a When the pressure in measurement mode increases above 320 mmHg the quick release valve is triggered and the pressure circuit opens When the press ure in test mode increases above 320 mmHg the measured value display flashes b Remove the plug from the cuff tube for testing Then attach the short con...

Страница 32: ...n test mode when the display indicates SYSTOLE DIASTOLE and PULSE 0 mmHg 5 Perform the usual tests for tolerance testing of the pressure reading leak testing of the pressure circuit take account of the cuff s settling time of one minute 6 Switch off the unit and restore the ori ginal connections C Security For security purposes the two halves of the unit can be joined together with a security mark...

Страница 33: ......

Страница 34: ......

Страница 35: ...BOSCH SOHN GmbH u Co KG Bahnhofstraße 64 D 72417 Jungingen T 49 0 74 77 92 75 0 F 49 0 74 77 10 21 E zentrale boso de Internet www boso de 07 2013 0124 ...

Отзывы: