25
6
Checking and fine-tuning measuring accuracy
!
Before commencing testing or fine-tuning work as described in Sections 6.2 or 6.3, the analyzer must be warmed up for a period
of 10 minutes. The cover of the device must be closed during this time.
It is essential to note the instructions in Section 6.1 before checking and fine-tuning measuring accuracy.
Please observe the propane factor of 0.5 for checking or fine-tuning the HC measuring channel.
Observe safety regulations when working with toxic gases.
6.1
Taking account of barometric altitude
Before checking or fine-tuning measuring accuracy, using a barometer read the current air pressure at the site in which testing or
adjusting is taking place.
In addition, find out what altitude is stored in the analyzer in input mode to Section 4.1.7.
If the analyzer is equipped with an air-pressure gauge (and this has been activated as described in Section 4.1.6), then check the
following:
- Does the altitude stored in the analyzer correspond to a mean air pressure which does not deviate by more than +50 mbar from the
currently measured air pressure? If there is a difference of 50 mbar or more, set the altitude in the analyzer to the current value.
If the analyzer is not equipped with an air-pressure gauge, the altitude in the analyzer must be set to the value which most accurately
corresponds to the current air pressure.
After you have finished checking or fine-tuning the device, set the altitude at which the device is to be used by the operator.
Altitude for mean air pressure
Altitude in m
Mean air pressure
above sea
level
in torr
in Hpa (mbar)
-700
825
1099
-600
815
1087
-500
806
1074
-400
797
1062
-300
787
1049
-200
778
1037
-100
769
1025
0
760
1013
100
751
1001
200
742
989
300
733
977
400
724
966
500
716
955
600
707
943
700
699
932
800
691
921
900
682
909
1000
674
899
1100
666
888
Altitude in m
Mean air pressure
above sea
level
in torr
in Hpa (mbar)
1200
658
877
1300
650
867
1400
642
856
1500
634
845
1600
627
836
1700
619
825
1800
611
815
1900
604
805
2000
596
795
2100
589
785
2200
582
776
2300
574
766
2400
567
756
2500
560
747
2600
553
737
2700
547
729
2800
539
719
2900
533
711
3000
526
702
6.2
Checking the indicating accuracy
The pump must be switched off. Select analysis mode with the "Select" button (press 10 x) and activate with the ”Enter
o
”
button.
Introduce a flow of test gas no. 4 into the device via the calibration-gas inlet.
Read the displayed values of the CO, CO
2
and HC measuring channels and compare with the certified test-gas values.
Maximum permitted indicating error = 5% from certified test-gas value.
If the value of one of the measuring channels exceeds the maximum permitted error, re-tune the measuring channel in ques-
tion to Section 6.3.
Содержание ETT 008.31
Страница 1: ...1 Repair instructions Exhaust gas analyzer ETT 008 31 ETT 008 38 ...
Страница 48: ...48 8 15 5 Sensor board component location diagram ...
Страница 51: ...51 8 17 1 Motherboard component location diagram ...
Страница 61: ...61 12 Diagrams circuit diagrams and gas route diagram 12 1 Circuit diagram 1 ...
Страница 62: ...62 12 2 Circuit diagram 2 ...
Страница 63: ...63 12 3 Gas route diagram ...
Страница 65: ...65 ...