Message
Description
EC10
The blood pressure monitor was not able to connect to your
telehealth appliance. Place your blood pressure monitor
closer to your telehealth appliance, and try again
EC20
The blood pressure monitor could not transmit your
measurements to your telehealth appliance. Place your blood
pressure monitor closer to your telehealth appliance, ensure
that telehealth appliance is on, and then try again.
EC87
The cuff did not inflate properly. Check the cuff hose for leaks
or loose connections, adjust the cuff position, and then try
again
EC91
The blood pressure monitor could not detect a stable
pressure signal. Try again
EC92
An error occurred with the internal hardware of the blood
pressure monitor. Try again
Robert Bosch Healthcare Systems, Inc
.
|
F.03D.601.424 Rev. A 12/2012
13 | en