173_58300000002379_ara_en_j.doc – 29.04.14
Seite 39 von 41
2.26.2 Sanitisation liquid
The liquid “HUWASAN med 2” is used for sanitisation. This solution
contains 1 per cent hydrogen peroxide.
The material number of the 5-litre canister is 00310550.
2.26.3 Water filter
After sanitisation is finished, insert a new water filter, if the appliance
has one.
2.26.4 Carrying out the sanitisation
D A N G E R
Exposed live parts
Risk of death due to electric shock
►
Disconnect the appliance from the mains.
►
Use the current-operated circuit breaker if live tests are
required.
►
Do not touch the housing, frame, or components.
►
Ensure that the resistance of the earthing conductor does not
exceed the standardized value.
C A U T I O N
Injuries may be caused by liquid spraying out
►
After finishing the sanitation work, empty the pressure tank
completely and depressurize it.
N O T E
Any microbiological substance in the water filter is not
sufficiently removed by sanitisation
►
Before sanitisation, replace the filter with a bypass.
►
After sanitisation, replace with a new filter.
N O T E
The microbiological substance is not sufficiently reduced in
the water system
To successfully sanitise the appliance, all quantity and time
specifications must be complied with.
The single sequences in detail are described in the specific repair
manual.
2.26.5
Maintenance instruction
C A U T I O N
Danger of mould formation!!
►
The pressure tank must be empty as well as all part must be
cleaned and dried after each use.
►
The pressure tank may be transported only pressure free!
►
Only dry and clean parts can be packed in the carrying case.