background image

Rev. 04-2010 USMAN2012-IT 

 

1 - 20 

 

 

 

 

OB OXIKIT PLUS REV11-2021_EN 

 

 

 

COMPLETE DISTRIBUTION OXYGEN KIT FOR AMBULANCE 

 

 

OB OXIKIT PLUS 

 

 

USER MANUAL

 

                   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PRODUCED BY

 

 

 

 

 

Содержание OB OXIKIT PLUS

Страница 1: ...Rev 04 2010 USMAN2012 IT 1 20 OB OXIKIT PLUS REV11 2021_EN COMPLETE DISTRIBUTION OXYGEN KIT FOR AMBULANCE OB OXIKIT PLUS USER MANUAL PRODUCED BY ...

Страница 2: ...Rev 04 2010 USMAN2012 IT 2 20 OB OXIKIT PLUS REV11 2021_EN ...

Страница 3: ...CTION HOSES 10 INSTALLATION OF THE OXYGEN BAR 10 LOW PRESSURE ELECTRIC TRANSDUCER 10 FUNCTIONAL CHECK ON THE EXCHANGER 11 CHECK ON PERMANENT CONNECTIONS 11 PREPARATION OPERATION 12 START 12 AFTER USE 12 LONG TERM NON USE 12 PREPARATION FOR RE USE 12 FUNCTIONAL CHECKS 13 PERIODIC CHECKS 13 CHECKING THE KIT FOR LEAKS 13 WHAT TO DO IN CASE OF ANY LEAK 13 REPLACEMENT OF LOW PRESSURE SEALING RINGS 14 T...

Страница 4: ...ices of Class 1 and higher European Directive 93 42 EEC Use the device only in the temperature range indicated Manufacturer Date of manufacture Materials used in the device may be recycled following appropriate procedure under national laws and local regulations Do not use if the packaging is not found to be intact Catalogue identification code Manufacturing lot number Serial number Use on the kit...

Страница 5: ...y prohibited to smoke or use an open flame near the device connected to the source oxygen cylinder whether in operation or with the valve closed Liquids must be prevented from entering the device as they may cause spontaneous combustion or explosion During assembly and disassembly of the pressure regulator it is prohibited to use any type of mechanical tool Screw the fitting to the oxygen cylinder...

Страница 6: ...th However since the use of oxygen in ambulances is linked to serious respiratory diseases in this case it is justified to use it for a longer time and this to safeguard the life of the patient Before proceeding with any type of operation on the kit it is mandatory to disconnect the source compressed oxygen cylinder Never place any parts of the device in water or fluid substances These substances ...

Страница 7: ...nality leakage and wear must be performed on the device At least every two years the filters of the inlet fitting must be replaced At least every five years the kit shall be fully overhauled and inspected by the manufacturer or an authorized service centre The hoses must be immediately replaced in case of leakage or damage and in any case every five years from the date of initial installation whet...

Страница 8: ...kit includes all components for installation on board ambulances It consists of an oxygen rail with two terminal points an exchanger and connecting hoses If required they are also available the pressure regulators type OXYWAY FIX III produced by WEINMANN in accordance with ISO 10524 1 The terminal points and pressure reducer outputs depend on customer requirements and applicable local specificatio...

Страница 9: ...elling plate the CE marking and the date of manufacture The manufacturer maintains an electronic system for the traceability of kit components and parts NOTE The CE marking on the device is to be understood to refer to the complete kit Separation or partial use of the components void the CE marking and certification The individual markings of components of the kit do not imply their conformity sep...

Страница 10: ... forces due to the vehicle s movement Consider these points carefully when installing the device in the ambulance INSTALLATION OF THE OXYGEN BAR The oxygen bar is made of aluminium alloy suitably treated to be compatible with oxygen It features a low pressure gauge terminals that are fixed on the front part of the bar itself and a manually controlled mechanical exchanger Fixing of the bar to walls...

Страница 11: ...essure regulators without opening them 2 Bring the exchanger to its central closed position 3 Slowly open the two cylinders by opening their valves Apply one manometer with high precision class at one of the terminal unit 4 Turn the knob of the exchanger to the position 1 and check the pressure value on the oxygen bar gauge Make a note of this value 5 Close the oxygen cylinders and wait 60 seconds...

Страница 12: ...mount of gas through to the kit For this reason at the end of each use the oxygen cylinders must always be closed at their valves and the exchanger knob set to its central closed position It is recommended to always empty the kit after use having already fully closed the cylinders LONG TERM NON USE If the device is not in use for prolonged periods of time the following precautions must be taken Av...

Страница 13: ...g rings for damage and or peeling of the chrome plating The safety valve located at the rear of the regulator must not be altered in any way Verify that terminals of the oxygen bar are undamaged and free of mechanical resistance to the insertion of medical device plugs Check the external condition if possible of the oxygen hoses 2 Check the sealing of the device see section Check on Permanent Conn...

Страница 14: ...rk environment is clean and use clean protective gloves Never clean or use detergents on the sealing rings Never use lubricants or greases to aid insertion of the rings into their lodgings TROUBLESHOOTING The table below summarizes possible faults defects and anomalies in the device and indicates the action to be taken by the operator in each case Fault defect anomaly Possible cause Remedy Leak be...

Страница 15: ...st be replaced immediately in case of any defect or leakage The exchanger is subject to mechanical wear and tear caused by the movement of the control knob Scheduled maintenance is required every five years after first installation and may only be performed by Oscar Boscarol Srl or one of its authorized service centres The following table shows the various maintenance frequencies of the kit Mainte...

Страница 16: ...al parts but do not alone indicate the conformity of the assembled kit The conformity of kits supplied by Oscar Boscarol Srl is declared following final testing and inspection of the complete device WARNING for the assembling of the pneumatic fittings see below you need a special tool available by our company Code Description Image SPS9128 5 pieces O ring ø 7x1 2 mm for hose fittings can be white ...

Страница 17: ... and in the technical data file for the kit The sizes of the pressure regulators are indicated in their respective instruction manuals Device specifications Regulator inlet fitting cylinder UNI 4406 DIN 477 N 9 NF E 29 650 F SS M10x1 Terminals on the oxygen bar UNI 9507 DIN 13260 or AFNOR NF S 90 116 Maximum kit intake pressure 4bar 1 0 400kPa 100 0 Minimum kit flow rate Qrv 100LPM Materials used ...

Страница 18: ...f medical devices that need oxygen flows exceeding this value inevitably determines variations in the nominal pressure of the system and may cause possible malfunctioning of the devices Symbols and identification of the device Symbols and labels are applied to all kit components to identify the device and date of manufacture The pressure regulators feature an indication of the due date of the next...

Страница 19: ...delivered with the product Detection by the assistance service of a fault and or defect attributable to material or production defects Absence of tampering modification or any other alteration not conforming to the original product as supplied In terms of safety reliability and functionality of the device Oscar Boscarol Srl may be held liable only if All interventions of assistance repair modifica...

Страница 20: ...Rev 04 2010 USMAN2012 IT 20 20 OB OXIKIT PLUS REV11 2021_EN Printed in Italy by Oscar Boscarol srl OB OXIKIT PLUS REV11 2021_EN ...