Emission test
Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Complies
Manufacturer instructions
– electromagnetic susceptibility
Immunity Tests
Compliance level
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV for contact discharge
± 15 kV for contact discharge
Radiated RF
IEC 61000-4-3
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3
3 V/m
80 MHz
– 2,7 GHz
80 % AM at 1 kHz
See Table 1
Fast electrical transients / burst
IEC 61000-4-4
±2 kV for power line
repetition frequency 100 kHz
Surge
IEC 61000-4-5
± 1 kV Line-to-line
± 2 kV Line-to-ground
Conducted RF
IEC 61000-4-6
3 V (0,15 MHz
– 80 MHz)
6 V in ISM bands between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
Power frequency (50/60 Hz) magnetic
field
IEC 61000-4-8
30 A/m
Voltage dips, short interruptions on
power supply input lines
IEC 61000-4-11
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
a 315°
0 % U
T
; 1 cycle and 70 % U
T
; 25/30
cycle
Single phase: at 0°
0 % U
T
; 250/300 cycle