4. If the electrode is contaminated with silver sulfide, immerse the electrode in the
thiourea solution to remove the deposit. After the electrode is processed,
recalibrate.
Note: Dispose of the cleaning solution or fill solution according to the handling
regulations for toxic or corrosive substances.