9) CHANNEL 1 CW ACQUISITION TESTS
a. Select internal calibrator and turn on at 0dBm:
Display the
Main Menu > Calibratr
menu to determine which calibrator is currently selected.
If the bottom line indicates “SelectInt”, then highlight that line and press Enter to select the
internal calibrator. If the bottom line indicates “SelectExt”, then the internal calibrator is
already selected.
Set
Main Menu > Calibratr > Int Level
to 0.0dBm.
Set
Main Menu > Calibratr > Int Signal
to “On”.
b. Plug in the CW sensor cable into the 4530 Sensor 1
connector and connect a model 51075 CW
sensor to the 4530 Calibrator.
Set
Main Menu > Chan 1 > MeasMode
to “CW”.
Set
Main Menu > Chan 1 > Params > dB Offset
to 0.00dB.
Set
Main Menu > Chan 1 > Params > Frequency
to 0.05GHz.
Set
Main Menu > Chan 1 > Params > Filter
to “Auto”.
Set
Main Menu > Chan 1 > Params > Range
to “Auto”.
c. Zero the sensor:
Press the Zero/CAL key, and press Enter.
Select “Zero Chan”. Press Enter.
The 4530 will perform a sensor fine zero adjustment operation.
Check for any error messages.
d. Perform a Fixed Calibration:
Press the Zero/CAL key.
Select “Fixed Cal”, and press Enter.
The 4530 will perform a fixed 0dBm calibration.
Check for any error messages, then press the Text key.
Channel 1 should display 0.000 dBm +/-0.002dBm.
e. Perform an Auto Calibration:
Press the Zero/CAL key.
Select “Auto Cal”. Press Enter.
The 4530 will perform an automatic sensor power sweep calibration.
Check for any error messages, then press the Text key.
Channel 1 should display 0.000 dBm +-0.002dBm.
10) CHANNEL 2 CW ACQUISITION TESTS (Model 4532 only)
a. Select internal calibrator and turn on at 0dBm:
Display the
Main Menu > Calibratr
menu to determine which calibrator is currently selected.
If the bottom line indicates “SelectInt”, then highlight that line and press Enter to select the
internal calibrator. If the bottom line indicates “SelectExt”, then the internal calibrator is
already selected.
Set
Main Menu > Calibratr > Int Level
to 0.0dBm.
Set
Main Menu > Calibratr > Int Signal
to “On”.
Scans by Artekmedia => 2009
Содержание 4530 Series
Страница 25: ...BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 24 Scans by Artekmedia 2009 ...
Страница 26: ...BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 25 Scans by ArtekMedia 2009 ...
Страница 27: ...BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 26 Scans by Artekmedia 2009 ...
Страница 29: ...BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 28 Scans by Artekmedia 2009 ...
Страница 31: ...BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 30 Scans by Artekmedia 2009 ...
Страница 32: ...BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 31 Scans by ArtekMedia 2009 ...