Bonso 840-10 Скачать руководство пользователя страница 2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Operating the Scale 

>    Make sure that the scale is plugged into the proper

outlet and on a flat and leveled surface   

>    Slide a proper display control unit   

( remote vs local display ) into the back of   

scale body, making sure the display control 

interface from adapter connects securely 

to adapter on scale   

 

Dual Weight Capacity 

 

      The P840 scale is a dual weight range scale 

      To change the weight capacity of the scale, 

  remove the round cover panel on the bottom of   

  the scale. Select desired weighing capacity by 

  sliding its weight selection control bar 

 

Key / Indicator 

 
 

ON 

TARE 

OFF

Power ON / OFF switch 

Press ONCE for weight tare   

MODE

Weight designator for 

 Kg, lb, lb:oz  

Operation 
 

A. Switching Weight Unit 

1. The measurement unit will be changed 

    from one unit to another ( kg, lb, lb:oz ) 

    by pressing MODE key   

2. The unit will always default in the last unit 

    used for its continuity 

     

B. Weighing (TARE) 

1. Place a (tare) item on the platter 

2. Press TARE key or press Foot-Tare button 

.    The scale displays “ 0 ” 

3. Place next (tare) item on the platter 

4. Press TARE key or press Foot-Tare button.   

The scale displays “ 0 ” 

5. Repeat step 3,4 until all tare items are 

weighted 

6. Place a weight item on the platter 

      The scale displays its weight 

7. Remove all items together 

      The scale displays “ – x.xx ” for its 

   total accumulated tare weight 

8. Press TARE key to clear its total weight 

   The scale returns to initial zero state “ 0 ” 

Calibration Procedure 

 

<<    The scale uses half weight of the   

total capacity to calibrate    >> 

1. With the scale turned ON , select the weighing 

      unit to be used for calibration by pressing 

   MODE key ( lb or kg ) 

2. Turn the scale OFF 

3. While holding down the MODE key with 

      nothing on the platter, turn the scale ON 

The scale displays “ cal

“ 

3. Press MODE key. The scale will briefly 

      display the weight unit to be used for   

      calibration ( kg or lb ), and then flash the 

   weight needed for calibration. 

4. Place the correct weight on the center of the 

   platform, and then press MODE key 

5. The scale will display CAL F, if the calibration 

      is successful and the process is finished 

6. The scale displays “CAL E” if un-successful 

    Repeat step 1 ~ 5 again for retry   

7. Turn the scale OFF, then turn ON to 

   return to Initial Zero state.  

<< NOTE >> 

      This procedure requires ONLY when it displays 

   erroneous weighing display with proper weight  

Options and Accessories 

Standard 

    1. Scale Platform 

        2. Weight display with a display control unit 

    Optional accessories 

        1. Remote display with a display control unit 

        2. Foot-Tare stepping unit   

Отзывы: