Страница 2: ...and Controls 6 7 Assembly and Use Non Portable Use 8 Installing the System 9 Operating Instructions 10 Oxygen Cylinder Duration 11 Care and Maintenance 12 Products Spare Parts Accessories 13 Troublesh...

Страница 3: ...ic oxygen requirements Statements in this manual preceded by the following words are of special significance WARNING Indicates there is a possibility of injury to you or others CAUTION Indicates there...

Страница 4: ...rated Avoid creation of any spark such as static electricity caused by any type of friction near oxygen equipment NOTE Oxygen will not burn however it does vigorously accelerate the burning of any fla...

Страница 5: ...ose the unit to extreme temperatures Always maintain a backup supply of oxygen Do not use humidifier bottles Do not use if leaking or damaged Refer repairs to authorized service personnel NOTE Oxygen...

Страница 6: ...ffects and eliminating unnecessary oxygen waste When we breathe approximately one third of the time is spent inhaling and two thirds exhaling As a result oxygen delivered by continuous flow is wasted...

Страница 7: ...auge that reads up to 4 000 PSI Selector Switch Enables the user to select the desired setting as well as OFF and cf 2 or cf 4 continuous flow When not in use the switch should be turned to the OFF po...

Страница 8: ...e the oxygen will be consumed at a much faster rate Return to another source before depleting the oxygen cylinder The continuous flow functions on the BONSAI VELOCITY conserver is factory preset at 2...

Страница 9: ...eal washer gasket An incorrect seal washer gasket may not be oxygen compatible or may cause an oxygen leak creating an increased fire risk DO NOT use the device if the manufacturer specified seal wash...

Страница 10: ...he corresponding holes on the cylinder post valve NOTE The cylinder adjustment knob should be aligned with the indentation on the post valve STEP 4 While holding the unit in place tighten the cylinder...

Страница 11: ...nula into position with the prongs in the nostrils and begin breathing The BONSAI VELOCITY conserver will now start to deliver oxygen The amount of oxygen delivered per pulse is determined by the sett...

Страница 12: ...0 times per minute as compared with ten 10 times per minute Please refer to the table below as a guide BONSAI SETTING 1 2 3 4 5 6 CF CF VELOCITY 2 lpm 4 lpm Cylinder Type Cylinder Volume Estimated Cyl...

Страница 13: ...e of the unit periodically with a clean lint free cloth Pay special attention to the oxygen inlet and outlet to make sure they remain free of dust etc If the oxygen inlet connection becomes contaminat...

Страница 14: ...t case of 50 NOTE Only manufacturer specified seal washers gaskets may be used with the BONSAI VELOCITY conserver These parts are available from your Home Care Provider STANDARD PRODUCT ORDER NUMBER D...

Страница 15: ...ylinder valve Vent hole is obstructed Turn the selector switch to the appropriate delivery setting Turn the selector switch to OFF wait a few moments then set at proper delivery setting Remove obstruc...

Страница 16: ...CLASSIFICATIONS The BONSAI VELOCITY conserver is classified as Class II per FDA 21 CFR Part 868 5905 Not suitable for use in the presence of flammable anesthetic mixture with air oxygen or nitrous oxi...

Страница 17: ...Dimensions Approximately 5 3 L 13 5 cm x 2 2 D 5 6 cm Weight Approximately 10 ounces 283 grams Operating Temperature 14 F to 104 F 10 C to 40 C Operating Relative Humidity 20 to 95 Operating Altitude...

Страница 18: ...service professionals shall immediately void this warranty This warranty does not extend to spare parts or accessories CHAD Therapeutics will not accept damages or charges for labor parts or expenses...

Страница 19: ...s Name _________________________________________________ Home Care Provider s Phone Number ________ ________________________________ Physician s Name _________________________________________________...

Страница 20: ...ww chadtherapeutics com United States Patent Number 6 484 721 PM 00078 13 B Printed in the U S A BONSAI and CHAD are registered trademarks of Inovo Inc Parent Company Drive Medical Design Manufacturin...