41
4. Electric current (4~20mA) Analog
Output (Option-5)
The BS-205 series of unit meters can output an analog signal for an indicated
value.(when the meter is equipped with an analog output unit)
The analog output of the BS-205 series allows for arbitrary output scaling.
This scaling can be achieved by setting the indication value for an output of the
maximum scale value(20mA for 4~20mA output range) in the dA.HI parameter of
the scaling data.
The current output occurs proportionally the current according to
the size of a weight In 4~20mA.
Analog output (0~10V, 4~20mA) option
Pin No.
Name
Description
19
V-OUT or A-OUT
Voltage output terminal (0~10V or 4~20mA)
20
COM
Common terminal for analog output.
▣ SPECIFICATION
output Voltage
4 ~ 20 mA DC Current out
Precision
Max 1/1000
Min Impedance
Under 500 Ω
Condition data (Analog output option)
Analog output LO
indication setup
0~99999
00000
Analog output HI
indication setup
0~99999
10000
Содержание BS-105
Страница 1: ...BS 105 Digital Indicator The Better Way forWeighing Measurements ...
Страница 27: ...26 4 2 3 2 4 2 2 3 2 4 4 4 ...
Страница 46: ...45 ...