26
Issue 010910
•
Disconnect equipment from the mains.
•
Unscrew the four fixing screws (12) of the
filter frame (11).
•
Remove the spacer bar and the filter
frame (11) from the hand-held lamp unit
(5).
•
Check the new filter unit for soiling.
•
If necessary, clean the new filter unit,
using a clean cloth and alcohol.
•
Place the filter frame on the casing.
•
Check that it is fitted correctly.
•
Put the spacer bar in position and fit the
four fixing screws. Do not over-tighten the
screws.
The UVAHAND 250 is ready for operation
again.
Cleaning
Cleaning the reflector, the UV filter and the UV lamp
If the equipment is operated in a dust-laden environment, the reflector, the UV filter and the
UV lamp become soiled very quickly.
The equipment must be opened in order to clean the reflector, the UV filter or the UV lamp.
Danger
Before opening the equipment, switch it off and disconnect from the
mains.
There is a risk of serious, even fatal injury!
Do not open the equipment until it has cooled down. If the equipment
has been in operation immediately before cleaning, let it cool for at
least 10 minutes before opening.
Acute danger of burns!
Cleaning
Cleaning the
reflector, the
UV filter and
the UV lamp
11
12 5
9