N2400048, REV 1 06/11
SPECIFICATIONS AND PERFORMANCE
Dimensions:
Accessory Heating
Accommodates
2 Block Model 113002
8” w x 9.8” d x 3.5” h
Blocks
113002-2
(20.3 x 24.9 x 8.6 cm)
110006
30 tubes up to 6mm
4 Block Model 113004
10.25” w x 11.25” d x 3.5” h
110010
20 tubes up to 10mm
113004-2
(26.0 x 28.6 x 8.9 cm)
110013
20 tubes up to 13mm
Net
Weight:
110016
6 tubes up to 25mm
2 Block Model 113002/113002-2
5 lbs. (2.3 kg)
110020
6 tubes up to 25mm
4 Block Model 113004/113004-2
6 lbs. (2.7 kg)
110025
30 tubes up to 6mm
Power:
110035
Combination:
2 Block Model 113002
115VAC, 200 watts
3 tubes up to 25mm
2 Block Model 113002-2
230 VAC, 200 watts
5 tubes up to 13mm
4 Block Model 113004
115VAC, 380 watts
6 tubes up to 6mm
4 Block Model 113004-2
230 VAC, 380 watts
110040
1.5 microcentrifuge
Temp Range:
a5°C to 120°C
tubes by 24 places
Uniformity:
±0.25°C @ 37°C
110045
0.5 ml tubes by 24 places
Stability:
±0.2°C from 37° to 100°C
110048
0.2 ml tubes by 48 places
Standard Block:
110051
96 well Microplate
Dimensions:
3” w x 3 ¾ ” d x 2” h
110096
0.2 ml Microtubes tubes
(7.6 x 9.5 x 5.1 cm)
110099
1.5 ml “dolphin” tubes 24 places
Shipping Weight:
2.0 lbs. (1.0 kg)
110002LID*
Lid for 110001 and
Net Weight:
1.0 lbs. (0.45 kg)
110002 Dry Baths
110004LID*
Lid for 110004 Dry Baths
*Temp Limited
Temperature Stability was determined by using NIST
traceable thermometers placed in thermometer hole in
standard 16MM aluminum blocks. Thermal grease was
used to ensure contact.
Temperature Uniformity was determined by using NIST
traceable thermometers placed in 16MM glass test tubes
with water in standard 16MM aluminum blocks. The
thermometers were suspended in the water. Six
positions were used in each block for the two-block
capacity unit and three positions were used in each
block for the four-block unit.
DEFINITIONS:
Uniformity
is the variation in temperature, at the
location described above, in the block at any instant in
time.
Stability
is the variation in temperature at a single
point over a long period of time.
Note:
The temperature in the test tube may be slightly
different than the temperature of the aluminum block.
The test tube material and the medium in the test tube
may affect the temperature.