DIGITAL DRY BATH INCUBATOR 

MODELS 113002, 113002-2, 113004, 113004-2 

(PLEASE READ THESE INSTRUCTIONS COMPLETELY BEFORE OPERATING)

 

 

PRODUCT INFORMATION 
 

Boekel Digital

 

Dry Bath Incubators are designed to be 

used with standard aluminum dry blocks (nominally 3” x 
3¾” x 2” high) of varying configurations for heating test 
tubes, microtubes and other small vessels. These 
Incubators are available in two sizes; Model 113002 has 
a two block capacity and Model 113004 has a four block 
capacity.  Both have an easily programmed PID 
(proportional integrative derivative) controller for rapid 
heat-up and accurate temperature control.  

The Boekel Dry Baths have recently been redesigned 
with a sleeker, more aesthetically pleasing appearance 
and an upgraded controller. The 113002 is a direct 
replacement for 111002 two block capacity unit. The 
113004 directly replaces the 111004. The Hot Start Dry 
Bath Model 111007  has been discontinued, but the 
113002 can be purchased with the new 110096 block 
(designed to hold a 96 well PCR plate) to provide a more 
versatile version of the Model 111007.  

UNPACKING 
 

1.  Remove the Incubator Base and Block Extractor 

from the carton. 

 

2.  Inspect for damage. Report all shipping damage 

to the carrier immediately. Shipping damages are 
covered by the carrier and repair and/or 
replacement for shipping damages must be 
coordinated through the carrier. 

 

3.  Retain all packaging material in the event that the 

unit must be returned. 

 

4.  Completely fill out and return the self-addressed 

Boekel Warranty Registration Card to Boekel 
Scientific.  If you prefer, you can register online at 
our website (www.boekelsci.com). 

 

SET-UP   
 

1.  Place  the  unit  in  a  clear,  draft  free  environment.                                          

Ensure that the surface on which the unit is 
placed will withstand the radiated heat produced 
by typical dry bath incubators. Ensure that the 
side ventilation holes are not blocked. 

 

2.  Plug the power cord into a grounded receptacle 

of appropriate voltage. 

 

3.  Place any standard block into the dry bath 

incubator. 

4.  When using the Dry Bath Incubator to heat    

microscope slides or 96-well plates, some form of 
well cover or lid (non-flammable) should be used. 

OPERATION 

1.  Turn the “Power switch ON (-). Ambient “Target 

Temperature" will be displayed on the digital 
readout. 

2.   The unit will automatically begin heating to the 

temperature last programmed in the unit. The unit 
retains its’ last programmed temperature in the 
event of accidental power loss or deliberate 
power down. 

 

SETTING TEMPERATURE 

 
The temperature controller has three buttons.  When the 
button on the left is depressed, it will display the “Target 
Temperature”.  When the left button is depressed 
simultaneously with the 

middle button, “Target 

Temperature” value is lowered.  When the left button is 
depressed simultaneously with the button on the right, 
the “Target Temperature” is increased.  When all buttons 
are released, actual block temperature is displayed.    
   
In the event of power loss, the Temperature Controller 
retains the last "Target  Temperature" value.  The 
flickering green dot in the upper left corner of the 
Controller Display is the heat indicator light.  
 
Allow time for temperature to re-stabilize after setting or 
resetting the temperature. 20 minutes should be allowed 
for block stabilization at high temperatures. 
 

CLEANING 
 

1.  Turn unit off and disengage power cord before 

performing any cleaning. 

 

2.  The well interior, top and sides of the unit can be 

cleaned using a damp sponge. Allow to dry 
thoroughly before restarting unit. 

 
SERVICE: RETURN AUTHORIZATION IS REQUIRED. 

Should service be required, contact your salesperson or 
call Boekel Scientific Customer Service at:

 

215-396-

8200. Units returned for repair must be in clean 
condition. 