13. TROUBLESHOOTING (2) 

 

PROBLEM POSSIBLE 

CAUSE 

SOLUTION 

LCD Display shows 

 “EE” 

The cuff was not properly 
applied or the rubber 
tube was bent or 
pressed. 

Review the cuff applying and 
testing sections of the instructions 
and re-test. 

The cuff position was not 
correct or it was not 
properly tightened 

Apply the cuff correctly and try 
again 

Body posture was not 
correct during testing 

Review the body posture and 
testing sections of the instructions 
and re-test. 

LCD Display shows 

abnormal result 

Speaking, arm or body 
movement, angry, 
excited or nervous during 
testing 

Re-test when calm and without 
speaking or moving during the test 

No response when you 

press button or load 

battery

 

Incorrect operation, or 
strong electromagnetic 
interference 

Take out batteries for five minutes, 
and then reinstall all batteries. 

 
SPECIFICATIONS 

 

1.  Product name: bodytalk Talking Upper Arm Blood Pressure Monitor KD-5962 
2.  Model: Upper Arm Blood Pressure Monitor KD-5962 
3. Classification: Internally powered, Type B applied part,IPX0,No AP or APG,Continuous 

operation 

4.  Machine size: approx.142mm×98mm×51mm (5 19/32” x 3 27/32” x 2”) 
5.  Cuff circumference: 22cm-30cm(8 21/32” -11 13/16”), 30cm-42cm(11 13/16” -16 17/32”) 

(Optional), 42cm-48cm(16 17/32” -18 29/32”) (Optional) 

6.  Weight: approx. 314g(15 1/16oz.)    (exclude batteries) 
7.  Measuring method: Oscillometric method, automatic inflation and measurement 
8.  Memory volume: 60 times with time and date stamp 

9.  Power source: DC:6V

600mA, batteries: 4 ×1.5V

SIZE AA   

10.  Range of cuff pressure: 0-295mmHg 
11.  Range of measurement: 30-280mmHg 
12.  Pressure accuracy: ±3mmHg 
13.  Pulse rate range: 40-180/min, accuracy: ±5% 
14.  Environmental temperature for operation: 5

℃~

40

(41

℉~

104

15.  Environmental humidity for operation: 

90%RH 

16. Environmental temperature for storage and transport: -20

℃~

55

(-4

℉~

131

17.  Environmental humidity for storage and transport: 

95%RH 

18.  Environmental pressure: 80kPa-105kPa 
19.  Battery life: Approx.2 months with new alkaline batteries and 3-min. usage per day. 

Please visit www.mybodytalk.com.au for more information and accessories supply. 

8

Содержание KD-5962

Страница 1: ...ATTERY LOADING 3 2 TOUCH SCREEN SWITCH 3 3 CLOCK AND DATE ADJUSTMENT 3 4 VOICE SETTING 4 5 CONNECTING THE CUFF TO THE MONITOR 4 6 APPLYING THE CUFF 4 7 BODY POSTURE DURING MEASUREMENT 5 8 TAKING YOUR...

Страница 2: ...night The highest value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while mos...

Страница 3: ...cally and non invasively The LCD display will show blood pressure and pulse rate The most recent 60 measurements can be stored in the memory with date and time stamp The voice function will ease the o...

Страница 4: ...to avoid unintentional operation 3 CLOCK AND DATE ADJUSTMENT a Once you install the battery or turn off the monitor it will enter Clock Mode and LCD will display time and date b While the monitor is...

Страница 5: ...and tighten it and close the Velcro fastener b Place the cuff around a bare arm 1 2cm above the elbow joint c While seated place palm upside in front of you on a flat surface such as a desk or table P...

Страница 6: ...are shown for self test You can check the LCD display according to the right picture Please contact the service center if segment is missing c If the monitor has stored results the LCD will momentaril...

Страница 7: ...ion is on the monitor will announce each result in the memory d When displaying the stored results the monitor will turn off automatically after 2 minutes of no operation You can also press the button...

Страница 8: ...es Arm or electronic blood pressure monitor was moved during testing Re test taking care to not move your arm or the electronic blood pressure monitor The cuff does not inflate properly or pressure fa...

Страница 9: ...Blood Pressure Monitor KD 5962 3 Classification Internally powered Type B applied part IPX0 No AP or APG Continuous operation 4 Machine size approx 142mm 98mm 51mm 5 19 32 x 3 27 32 x 2 5 Cuff circumf...

Страница 10: ...2006 for details of verification method if you need 9 Information regarding potential electromagnetic or other interference between the blood pressure monitor and other devices together with advice re...

Страница 11: ...more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from t...

Страница 12: ...ions or other information which will assist the user s appropriately qualified technical personnel to repair those parts of equipment which are designated repairably can be supplied by us 9 The monito...

Отзывы: