Should any malfunctions occur while using the TENS, check 

-  whether the switch/control is set to the appropriate form of therapy. Adjust the 

control correctly.

-  whether the cable is correctly connected to the device. The 

cables should be inserted completely into the sockets.

-  whether the impulse display led is illuminated. If necessary, insert 

a new battery.

-  for possible damage to the cable. Change the cable if any damage is detected.

* If there is any other problem, please return the device to your distributor. Do 
not try to repair a defective device.

The ZZAN602 devices are in compliance with the EN 60601-1-2:2001 and EN 
60601-1:1990+A1:1993+A2:1995+A13: 1996 safety standards.

All ZZAN602 models carry a warranty of one year from the date of delivery. The 
warranty applies to the stimulator only and covers both parts and labor relating 
thereto. The warranty does not apply to damage resulting from failure to follow 
the operating instructions, accidents, abuse, alteration or disassembly by 
unauthorized personnel.

CHAPTER 21 : MALFUNCTIONS

CHAPTER 22: CONFORMITY TO SAFETY STANDARDS

CHAPTER 23 : WARRANTY

Содержание ZZAN602

Страница 1: ...ANALOG TENS UNIT Instruction Manual REORDER NO ZZAN602...

Страница 2: ...odes Lead Wires 10 12 Lead Wire Maintenance 10 13 Electrode Options 11 14 Electrode Placement 11 15 Tips For Skin Care 13 16 Application of Re usable Self Adhesive Electrodes 14 17 Adjusting the Contr...

Страница 3: ...ation is a non invasive drug free method of controlling pain TENS uses tiny electrical impulses sent through the skin to nerves to modify your pain perception TENS does not cure any physiological prob...

Страница 4: ...le allergic to gel skin irritation and electrode burn are potential adverse reactions 4 Read operation manual before use of TENS 5 We emphasize that patient with an implanted electronic device for exa...

Страница 5: ...n of pain which would otherwise serve as a protective mechanism 7 Electronic monitoring equipment such as ECG monitors and ECG alarms may not operate properly when TENS stimulation is in use 8 There s...

Страница 6: ...ce is provided with two controllable output channels each independent of each other An electrode pair can be connected to each output channel The electronics of the ZZAN602 create electrical impulses...

Страница 7: ...period of 4 seconds from the control setting value to a value which is 40 less The lower pulse rate will continue for 1 second Then increase linearly over a 4 seconds period to its original value The...

Страница 8: ...ECTRODES LEADS 02 ELECTRODES 03 9V BATTERY TYPE 6F22 04 BELT CLIP 05 BATTERY CASE COVER 06 LEAD CONNECTOR 07 MAIN PCB 08 INTENSITY KNOB 09 B N M SWITCH 10 PULSE WIDTH KNOB 11 PULSE RATE KNOB Each set...

Страница 9: ...fibers The choice of which pulse duration to use is partially dependent upon the Treatment Mode and Protocol selected refer to the appropriate section PULSE RATE The Pulse Rate hertz or pulses per sec...

Страница 10: ...tion of Conventional TENS and Low Rate TENS In Burst Mode the treatment frequency is fixed by the instrument and is not adjustable with the Frequency Rate control Modulated Mode attempts to prevent ne...

Страница 11: ...s see drawing one or two sets of wires may be used After connecting the wires to the stimulator attach each wire to an electrode Use care when you plug and unplug the wires Jerking the wire instead of...

Страница 12: ...here If the initial results are not positive feel free to experiment Once an acceptable placement has been achieved mark down the electrodes sites and the settings on the patient s reference sheet of...

Страница 13: ...linician has recommended Let this dry Apply electrodes as directed 4 Many skin problems arise from the pulling stress from adhesive patches that are excessively stretched across the skin during applic...

Страница 14: ...ectrodes on the liner and remove the lead wire by twisting and pulling at the same time Care and Storage 1 Between uses store the electrodes in the resealed bag in a cool dry place 2 It may be helpful...

Страница 15: ...to a frequency of approximately 30 Hz At higher frequencies the lamp will appear to be constantly illuminated 3 On Off Switch and Intensity Control If both controls are in the off position white mark...

Страница 16: ...gh the skin each second By turning these controls the number of current impulses per second Hz for both channels can be continually adjusted Unless otherwise instructed turn the pulse rate control to...

Страница 17: ...and remove 3 Remove the battery from the compartment 4 Insert the battery into the compartment Note the polarity indicated on the battery and in the compartment 5 Replace the battery compartment cover...

Страница 18: ...acturer s recommended charging time has been completed unplug the charger and remove the battery 4 Batteries should always be stored in a fully charged state To ensure optimum battery performance foll...

Страница 19: ...e device and accessories in carrying box and keep it in cool dry place 6 The packed TENS device should be stored and transported under the temperature range of 20 60 relative humidity 20 95 atmosphere...

Страница 20: ...if any damage is detected If there is any other problem please return the device to your distributor Do not try to repair a defective device The ZZAN602 devices are in compliance with the EN 60601 1 2...

Отзывы: