26
P063GUI001 - User guide eCelsius Performance Connect.1 - EN
5. TECHNICAL DATA
Dimensions:
Lenght: 17.7 mm
Diameter: 8.9 mm
Weight: ≈ 1.7 g.
Temperature range:
25°C - 45°C
Accuracy of temperature sensor:
± 0.1°C in the range 36-41°C (physiologic range), ± 0.13°C out of the
physiologic range
Température resolution:
0.01°C
Capsule memory capacity:
the 2000 last temeprature values are stored into the capsule internal memory
Maximal transmission distance with ePerf Connect:
around 1 to 3m (environment depending)
Power:
Autonomous system inclding 4 zinc-oxide batteries
Autonomy:
20 days for a 30sec measurement period
Communication frequency :
ISM Band 433MHz - 434MHz
Measurement period available
: 15s - 30s - 1min - 2 min - 5min
Plastic :
Biocompatible PVC
The manufacturing date is indicated on the blister of the pill. The shelf life of the eCelsius Performance
capsule is validated for a use up to 24 months after this date. Beyond that date, device performances and
autonomy are not guaranteed.
ENVIRONMENTAL CONDITIONS
In operation
- Temperatures in the range 25-45°C
- Ingress protection (IP): X8 (Material supportanting prolonged immersion)
Storage and/or transportation
- Humidity between 30 and 80% of relative humidity
- Atmospheric pressure between 700hPa and 1060hPa
- Ambiant temperature between 5 and 35°C
- Avoid sprayed water
- Avoid exposure to sunlight
eCELSIUS PERFORMANCE CAPSULE CHARACTERISTICS