6.10 USAGE OF ELECTRODE PADS
a. If you want to reposition the electrodes during the
application, turn off the device first.
b. The use of the electrodes may lead to skin irritations.
If you experience such skin irritations, e.g. redness,
blistering or itching, discontinue using them. Do not use
the unit permanently on the same body part, as this
may also lead to skin irritation.
c. Self-adhesive pads are for single patient use only.
d The electrode must connect entirely to the skin surface
to prevent hot spots, which may lead to skin burns.
e. Do not use the electrode pads more than approx. 10
times, as connection between the electrodes and the
skin deteriorates over time.
f. The adhesive force of the electrodes depends on the
user’s skin properties, storage conditions, and the
number of applications. If your electrode pads no longer
fully stick to the skin’s surface, replace them with new
ones. Stick the electrode pads back onto the protective
film after use and store them in the storage bag to
prevent them from drying out. This retains the adhesive
force for a longer period.
Caution:
1) Before applying the electrodes, it is recommended for
users to wash their skin and then dry it.
2) Never remove the electrodes from the skin while the
device is still on.
3) Only use the electrode pads provided by the
manufacturer or authorised distributor. Usage of other
companies’ products could result in injuries to the user.