![Body Clock Health Care Elle TENS Скачать руководство пользователя страница 33](http://html.mh-extra.com/html/body-clock-health-care/elle-tens/elle-tens_instructions-for-use-manual_2782762033.webp)
Programme
Display
Output Frequency / Duration
Contstant
CON-
STANT
Pulse Frequency 100Hz
Pulse Width 20-250
애
s
Burst
BURST
32Hz-2 bursts per second, 14 pulses per burst
Pulse Frequency 2Hz (each Hz = 14 pulses per burst)
Pulse Width 20-250
애
s
Massage
MAS-
SAGE
Pulse Frequency 100Hz fixed frequency
modulating on pulse width
Pulse Width 20-250
애
s
Mixed
MIX
3 seconds in mode 1, 3 seconds in mode 2, repeat
Pulse Frequency 2Hz (each Hz = 16 pulses per burst)
and 100Hz (repeated)
Pulse Width 20-250
애
s
Timer
min
(C) Continuous, 15 Minutes, 30 Minutes
45 Minutes and 60 Minutes
Output Channel
Dual Channel
Output Waveform
Asymmetrical bi-phasic rectangular
Output Voltage
0-60V adjustable in 15 steps (500Ω load, peak to 0)
Output Intensity
0 - 120mA adjustable in 15 steps (500Ω load). Intensity
level will drop back to 0 after mode changes
PAIN mode
Programme
Display
Output Frequency / Duration
Urge
PROG 1
5Hz / 3 mins 10Hz / 5 mins 15Hz / 5 mins
20Hz / 5--Hz / 5 mins
10Hz / 5 mins 5Hz / 2 mins
Total time = 30 mins*
Stress
PROG 2
10Hz / 5 mins 35Hz / 5 mins 50Hz / 10 mins
35Hz / 5 mins 10Hz / 5 mins
Total time = 30 mins*
Mixed
PROG 3
10Hz / 5 mins 20Hz / 5 mins 30Hz / 5 mins
40Hz / 5 mins 50Hz / 5 mins 40Hz / 5 mins
30Hz / 5 mins 20Hz / 5 mins 10Hz / 5 mins
Total time = 45 mins*
After Care
PROG 4
Frequency Modulation
Frequency increases from 2 Hz to 10 Hz in 4 secs.
and then back to 2Hz in another 4 secs.
Total cycle = 8 secs
Total time = 45 mins*
Pulse Duration
Pulse width fixed at 200
애
s for all progammes
Output Channel
Single Channel
Output Waveform
Bi-phasic rectangular
Output Voltage
0-40V adjustable in 14 steps across a 500 ohm load
Output Intensity
0-80mA adjustable in 14 steps across a 500 ohm load
PELVIC mode
33
* Timer countdown displayed on screen.
Содержание Elle TENS