background image

HF-MS System Operating Instructions 

page 1 of 6

 

PC Interface Socket

Back

Antenna Socket

Status Buzzer

Status LEDs

Front

BOBE HF-MS System 

 

HF-MS-M (receiver) 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
HF-MS-Sx (sender) 
 

    e.g. HF-MS-S1 for  

 

 

e.g. HF-MS-S6 for 

    Mitutoyo Digimatic caliper 

 

or

 

default OptoRS232 measuring device 

 
 
 

  

 

 

 
 
 
 

General Information: 

 

• 

The  Bobe  HF-MS  System  provides  computer  aided,  cable-free  measuring  by  replacing  the  cable 
connection  between  the  measuring  device  and  the  interface  box  or  PC  by  radio  transmission. 
Depending  on  the  environmental  conditions  you  can  bridge  a  maximum  of  80m  in  the  open  air  and 
approx. 10m-15m in a hall. 

• 

The system currently exists for all measuring devices with a Digimatic interface and default OptoRS232 
interface and for the Mahr EX 16 caliper or compatible measuring devices (for devices with an RS232C 
interface  upon  request).  You  can  operate  a  maximum  of  99  measuring  devices  concurrently  in  a 
system.  A  number  is  allocated  to  each  measuring  device  allowing  you  to  identify  the  individual 
measured  values  correctly.  (With  receiver  version  1.xx  and  sender  version  1.xx,  2.xx  or 3.xx you can 
operate only eight measuring devices.) 

• 

A  system  consists  of  a 

receiver

  (HF-MS-M)  which  is  connected  directly  to  the  PC,  and  at  least  one 

sender 

(HF-MS-Sx)  which  is  connected  directly  to  the  measuring  device.  The  transfer  of  measured 

values  is  triggered  by  pressing  the  data  button  on  the  measuring  device  or  sender  (depending  on  the 
measuring device type). The transfer of measured values from the receiver to the PC can be set by an 
interface  with  MUX10,  MUX50  or  Bobe  command  set  and  has  to  be  evaluated  correctly  by  the  PC 
software.  

• 

The  radio  frequency  is  set  to  either  (1)  868,300MHz,  (2)  869,750MHz,  (3)  869,2625MHz,  (4) 
869,800MHz  or  (5)  869,925MHz.  The  type  plate  shows  what  frequency  your  system  uses.  Make  sure 
that  only  devices  with  the  same  frequency  work  together  and  that  no  foreign  device  sends  signals  on 
this frequency in a vicinity of 300m. Otherwise the data transfer might be faulty. 

Data Button

Connector variants HF-MS-Sx: (x stands for the 
number) 

1. Mitutoyo Digimatic caliper 
2. Mitutoyo Digimatic Micrometer 
3. Mitutoyo Digimatic IDF Indicators 
4. Mahr - for EX series measuring devices 
5. TESA - for capa µ series measuring devices 
6. OptoRS232 from Sylvac, Mahr, Helios, Preisser etc. 
7. Mecmesin AFG xxxN (20 pins) 
8. Mecmesin AFG xxxN (15 pins) 
9. Mahr Height Measuring Instrument CX2 
10. Mitutoyo Digimatic caliper IP65 
11. Mahr Extramess 2000/2001 

Содержание HF-MS System

Страница 1: ...uring devices A system consists of a receiver HF MS M which is connected directly to the PC and at least one sender HF MS Sx which is connected directly to the measuring device The transfer of measured values is triggered by pressing the data button on the measuring device or sender depending on the measuring device type The transfer of measured values from the receiver to the PC can be set by an ...

Страница 2: ...n also fix the HF box to the wall to achieve a higher location In any case the distance between the receiver and sender must not exceed 80m in the open air and approx 10 15m in the hall In separate rooms the distance might have to be shorter Step 2 Connect the receiver HF MS M to the mains and PC Use the USB cable supplied to connect the receiver HF MS M to a free USB port of your PC The receiver ...

Страница 3: ...ta button on your measuring device with default OptoRS232 you have to press the button on the sender to transfer the value If need be read the operating instructions of your measuring device When the receiver receives the value correctly two tones sound and the value is forwarded to the PC Only press the data button again once the measured value has been transferred to the PC and no tone can be he...

Страница 4: ...nt device was received Other devices which are not used for radio transfer might also interfere Try to localize and remove these devices If the receiver issues two tones the value was received correctly If it does not appear in the PC software the connection between receiver and PC is faulty or the PC software has been configured incorrectly Operation with Several Measuring Devices To operate seve...

Страница 5: ...ry the right way round and press the data button on the measuring device or sender within 2 seconds This will save the two digits before and after or both digits after the point as measuring device number As it is the case every time you insert a battery a measured value will not be transferred but the version number of the sender firmware will be transferred as a measured value The computer will ...

Страница 6: ...mm x 75mm x 45mm Weight approx 300g Sender Operating Frequency 1 868 300MHz 2 869 750MHz 3 869 2625MHz 4 869 800MHz or 5 869 925MHz Transmitting Power 7dBm 7mW Range up to 80m in the open air min 10 15m in the hall License According to the official decree of the BaPT 222 ZV 125 and the European licensing standard ETS 300 220 the device has a general license and doesn t require any further license ...

Отзывы: