41
Clean the smart device
Dampen a clean, non‐abrasive, cotton cloth
with an ethanolic‐based liquid cleaner, and
wipe the screen clean.
Apply the cleaner to the cloth rather than the
surface of the screen.
Clean and disinfect system surfaces
1 Turn off the system.
2 Disconnect the system from the power supply,
or remove it from the stand.
3 Clean the exterior surfaces using a soft cloth
lightly dampened in a mild soap or detergent
cleaning solution to remove any particulate
matter or body fluids.
Apply the solution to the cloth rather than the
surface.
4 Mix the disinfectant solution compatible with
The system, following disinfectant
label instructions for solution strengths and
disinfectant contact duration.
5 Wipe surfaces with the disinfectant solution.
6 Air dry or towel dry with a clean cloth.
To disinfect the transducer and its cable, use the
immersion method or the wipe method.
Immersible transducers can be disinfected only if
the product labeling indicates they can be used
with an immersion method.
Clean and disinfect a transducer
1 Remove any transducer sheath.
2 Clean the surface using a soft cloth lightly
dampened in a mild soap or detergent
cleaning solution to remove any particulate
matter or body fluids.
Apply the solution to the cloth rather than the
surface.
3 Rinse with water or wipe with
water‐dampened cloth, and then wipe with a
dry cloth.
4 Mix the disinfectant solution compatible with
the transducer, following disinfectant label
instructions for solution strengths and
disinfectant contact duration.
5 Wipe surfaces with the disinfectant solution.
6 Air dry or towel dry with a clean cloth.
7 Examine the transducer and cable for damage
such as cracks, splitting, or fluid leaks.
If damage is evident, discontinue use of the
transducer, and contact BMV or your local
representative.
Caution:
Transducers must be cleaned after
every use. Cleaning transducers is
necessary prior to effective
disinfection. Ensure that you follow
the manufacturer's instructions
when using disinfectants.
Using a non‐recommended
cleaning or disinfection solution,
using incorrect solution strength, or
immersing a transducer deeper or
for a longer period of time than
recommended can damage or
discolor the transducer and void
the transducer warranty.
Do not allow disinfectant to contact
metal surfaces. Use a soft cloth
lightly dampened in a mild soap or
compatible cleaning solution to
remove any disinfectant that
remains on metal surfaces.
Caution:
Attempting to disinfect a
transducer or transducer cable
using a method other than the one
included here can damage the
transducer and void the warranty.