3.4
AMS-K-THK
The AMS-K-THK cassette can be used for temperature measurements with thermocouple
(type K).
3.5
AMS-K-BLANK
To protect the AMS system against external influences, unused slots can be covered with
the AMS-K-BLANK, a blank panel without a connector.
4
Slots for thermocouples or current shunts
If the cassette is equipped with a thermocouple amplifier, the cold junction compensation element delivered
with the module must be installed on the cassette (J2). If the cassette is equipped with a current amplifier, the
shunt delivered with the module must be installed on the cassette (J1). The positions of the slots J1 and J2 are
illustrated in Figure 1.
5B modules from bmcm do not need any extra equipping. These options have been implemented in the mod-
ules already.
5
Important notes for using the AMS-K
•
The AMS-K cassettes are only suitable for extra-low voltages - please observe the relevant regulations!
•
Only use an electrical isolated power supply unit (with CE).
•
All accessible pins are electrostatic devices. Workplace must be conductive during installation.
•
Only use non-solvent detergents for cleaning. The product is designed to be maintenance-free.
•
The product must not be used for safety-relevant tasks. With the use of the product, the customer becomes manu-
facturer by law and is therefore fully responsible for the proper installation and use of the product. In the case of
improper use and/or unauthorized interference, our warranty ceases and any warranty claim is excluded.
Do not dispose of the product in the domestic waste or at any waste collection places. It has to be either duly dis-
posed according to the WEEE directive or can be returned to bmcm at your own expense.
6
Technical data
(typical at 20°C, after 5min., +5V supply)
•
Electrical data
Power supply (from AMS system):
+5V DC ±5%, max. 0.2A DC
Galvanic isolation:
depending on the module in use
Max. permissible potentials:
max. ±60V DC (VDE)
•
General data
Signal connection:
at the AMS-K cassette panels (AMS-K-BIN5: 5-pole Binder female of the 712 series; AMS-K-BNC:
BNC female; AMS-K-CO5: 5-pole terminal connector; AMS-K-THK: thermocouple conn. (type K)
CE standards:
EN61000-6-1, EN61000-6-3, EN61010-1; for decl. of conformity (PDF) visit
RoHS and WEEE compliant // WEEE Reg.-No. DE75472248
Protection:
IP20
Temperature ranges:
operating temp. -25..50
°
C, storage temp. -25..70
°
C
Relative humidity
0-90% (not condensing)
Dimensions (without connector):
W x H x D: 20mm x 128mm x 60mm
Delivery:
device, description
Accessories:
AMS-K-CO5: 5-pole screw-type terminal blocks ZU5ST; 5-pole Binder female ZU5S712
Warranty:
2 years from date of purchase at bmcm, claims for damages resulting from improper use excluded
Manufacturer: BMC Messsysteme GmbH. Subject to change due to technical improvements. Errors and printing errors excepted. Rev. 4.0