BMB MEDICAL – CAREXIA – 1VCRX092020
79
4.2
TRANSPORT
AND
STORAGE
Storage and transport conditions:
Ambient conditions
Operation
Ambient temperature
-10 to +50 °C
Relative humidity
20 to 90% at 30 °C - no condensation
Atmospheric pressure
700 to 1060 ha
4.2.1
TRANSPORT
During chair transport with a view to its storage, the following is recommended to prevent any
impacts or falls:
-
Strap it to a surface of the means of transport
-
Cover it
-
Engage the brakes
-
Set it to its low position
Damage caused during transport is not covered by the warranty. The costs of repair and
replacement of damaged parts will be borne by the customer.
4.2.2
STORAGE
When the chair is in storage, ensure that:
-
Its brakes are engaged
-
It is covered
The product should be stored in a clean, ventilated, temperate and dry place, protected from
the elements and from direct sunlight.
During storage, no heavy weights should be placed on the chair.
The chair must not be considered as a support surface for any other type of equipment.
The device must not be exposed to, or even less be in contact with, sources of heat, combustion
and flammable agents.
In the event of prolonged storage, the battery should be recharged every 6 months,
proceeding as follows:
-
Connect the mains power cord.
-
Briefly press one of the remote control functions to initiate charging.
-
Allow the battery to charge for 8 hours.
Содержание CAREXIA FP
Страница 1: ...BMB MEDICAL CAREXIA 1VCRX092020 1 CAREXIA MANUEL D UTILISATION USER MANUAL FRAN AIS ENGLISH...
Страница 7: ...BMB MEDICAL CAREXIA 1VCRX092020 7 VUE D ENSEMBLE...
Страница 9: ...BMB MEDICAL CAREXIA 1VCRX092020 9 VUE D ENSEMBLE...
Страница 11: ...BMB MEDICAL CAREXIA 1VCRX092020 11 VUE D ENSEMBLE...
Страница 13: ...BMB MEDICAL CAREXIA 1VCRX092020 13 VUE D ENSEMBLE...