BMB MEDICAL – CAREXIA – 1VCRX092020
75
-
Tell the patient not to place his/her hands between the seat and backrest to avoid any
pinching risk.
-
To ensure correct use of the product, only proceed if the weight is correctly distributed
over the device.
-
The “variable height” function should be used when handling the patient in order to
work at an ergonomic height.
-
Once movement is no longer the required, the chair should be immobilized with the
brakes. Next, check that the wheels are locked by attempting to move the chair. Failure
to observe this manipulation could lead to the fall of a person leaning against the
chair.
-
Do not use the MD on a floor with a slope of more than 10%.
-
To avoid the risk of patient or user injury of IV pole damage when moving the chair,
ensure that the IV pole is firmly secured.
-
To avoid damaging the equipment during transport, ensure that the IV pole passes
safely through all door frames and under all light fittings.
-
Never use the chair in the presence of an anesthesia mixture flammable with air, with
oxygen, or with nitrous oxide as this could cause a fire and/or explosion.
-
To prevent failure, electromedical equipment requires special EMC-related precautions,
described
-
Do not manually raise the legrest.
-
Work on electrical parts should only be performed by qualified and authorized staff.
-
Used to accommodate a person of small stature (less than 1.30 m tall), or under the
age of 12 years.
-
Do not alter the chair. Any changes made to the system could lead to unpredictable
operation that could in turn jeopardize the safety of the user or patient. This would also
void the chair’s warranty.
-
The chassis cover load capacity is of 35 kg. Do not sit or stand on the cover.
-
To avoid material damage or injury, do not use the IV pole to pull or push the product.
-
To avoid damaging the equipment, the maximum permissible load of the IV pole is
15 kg.
-
Use by a person who has not first familiarized themselves with this user manual.
-
Use by any non-caregiver who has not received the authorization and explanations
from a qualified individual.
-
Use of the chair with a load that exceeds the safe working load (see technical
characteristics).
-
Use outside of the establishment, or on unprepared ground.
-
Washing with large amounts of water, with a high-pressure jet or with steam and
passage through a wash tunnel.
-
Connection, by any means, of an electric device to the chair.
-
Any use that does not fall within the operating context of a hospital chair.
All manipulations relating to the aforementioned precautions for use are described in
paragraph II - Instructions for use
Содержание CAREXIA FP
Страница 1: ...BMB MEDICAL CAREXIA 1VCRX092020 1 CAREXIA MANUEL D UTILISATION USER MANUAL FRAN AIS ENGLISH...
Страница 7: ...BMB MEDICAL CAREXIA 1VCRX092020 7 VUE D ENSEMBLE...
Страница 9: ...BMB MEDICAL CAREXIA 1VCRX092020 9 VUE D ENSEMBLE...
Страница 11: ...BMB MEDICAL CAREXIA 1VCRX092020 11 VUE D ENSEMBLE...
Страница 13: ...BMB MEDICAL CAREXIA 1VCRX092020 13 VUE D ENSEMBLE...