5.0    Error messages

The following error messages appear for the following reasons.

4.0    Calibration

Calibration of conductivity is not required for this unit as it is factory calibrated. 
However; if you wish to calibrate the unit follow the instructions below.

YOU MUST CLEAN THE PROBE BEFORE 

CALIBRATING.

See section 2.0.

Rinse probe in fresh water and place it in a known 

standard solution. See chart below for the correct 

solution.

Wait for reading to stabilize.

Hold down the cal button for 3 seconds until CAL 

appears. 

Release button and CAL P should be displayed. If Err 

is displayed check the probe is clean and that the 

calibration solution is fresh and uncontaminated.

A check mark will appear on the screen to indicate that 

the calibration was successful. The check mark will 

disappear after 30 days. To reset back to factory default 

remove/replace battery.

NOTE: If you need to test or calibrate in a 1500 ppm 

solution, you MUST set the pen to EC, then multiply your 

result by 540. If calibrating, multiply 2.8 by 540 

(2.8 x 540 = 1512).
This unit DOES NOT measure in the 540 ppm scale.

Temperature 

over range

ppm over range

Temperature error

Hardware error

Temperature 

under range

English

5

1

2

3

4

EC

ppm 500

(TDS)

ppm 700

(EC x 700)

Solution value

2.77

1385

1940

Displayed value

2.8

1390

1940

1385

 ppm

Hold for 3

seconds

Содержание ppm pen

Страница 1: ...nt 4 Calibration 5 Error messages 5 Troubleshooting guide 6 Technical specifications 6 Information about the scales available on the Bluelab ppm Pen 7 Product guarantee 8 Blulelab Probe Care Kits 9 Co...

Страница 2: ...press to turn off Calibrate button See calibration section Units button Hold until units flash then short press to change units Screen will change back when no buttons have been pressed for 3 seconds...

Страница 3: ...wer button again To change units Hold down the units button for 3 seconds until the conductivity and temperature units start flashing Short press units button again to cycle between unit combinations...

Страница 4: ...orms a film on the probe face with no beads of water If beading is present repeat the cleaning process Replace shroud and test in a known solution to ensure the unit has been adequately cleaned 3 0 Ba...

Страница 5: ...layed If Err is displayed check the probe is clean and that the calibration solution is fresh and uncontaminated A check mark will appear on the screen to indicate that the calibration was successful...

Страница 6: ...ecifications Range 0 0 10 0 EC 0 7000 ppm 700 ppm 0 5000 ppm 500 ppm TDS 0 50 C 32 122 F Resolution 0 1 EC 10 ppm 700 10 ppm 500 1 C 1 F Accuracy 0 1EC 25 C 2 77EC 50 ppm ppm 500 25 C 1385 ppm 70 ppm...

Страница 7: ...rent reasons Did you even know there are more than two scales The most widely used scales in Hydroponics are the 500 scale 650 scale and the 700 scale What s the difference The ppm 500 scale is based...

Страница 8: ...incorrectly inserted or damaged battery contacts or connections caused by incorrect battery replacement or ingress of moisture from incorrect positioning of the battery cap and waterproof seal NO RES...

Страница 9: ...in the instrument s brain or micro computer and a reading is displayed If the information sent back from the probe is inaccurate due to probe surface contamination then the reading will be inaccurate...

Отзывы: