IntraOs 70  

X-ray Equipment

 

 

Service & Installation Manual

 

Blue X Imaging Srl Via Idiomi 1/8-33 20090 Assago ITALY e-mail [email protected]

 

Содержание IntraOs 70

Страница 1: ...IntraOs 70 X ray Equipment Service Installation Manual Blue X Imaging Srl Via Idiomi 1 8 33 20090 Assago ITALY e mail bluex bluex it...

Страница 2: ...8 Patient size large 9 Patient size small 10 Maxillary incisor 11 Maxillary canine or premolar 12 Maxillary molar 13 Mandibular incisor 14 Mandibular canine or premolar 15 Mandibular molar 16 Bite wi...

Страница 3: ...x 20 POWER SUPPLY 24 V LED FUSE 3 T 200 mA 5 x 20 OUTPUT LINE TO TUBEHEAD INPUT MAINS LINE NEUTRAL LIVE JUMPER FUSE 2 MAINS SWITCH FUSE 1 FUSE 2 T8A 6 3x32 Layout Power Board Timer AutoSet 3 52 POWER...

Страница 4: ...uex it IntraOs 70 Dental X ray Equipment Service Installation Manual English Edition Version 4 0 March 2008 Printed 30 05 2008 12 11 00 Code 69 500 00210 Blue X Imaging is committed to Total Quality T...

Страница 5: ...ENANCE 4 1 Maintenance of the Wall Support 4 2 Maintenance of the Extension Arm 4 3 Maintenance of the Folding Arm 4 4 Maintenance of the Tube Head 4 5 Maintenance of the Beam Limiting Device 4 6 Main...

Страница 6: ...requirements see section 3 1 3 at page 13 To connect the timer see section 3 1 10 at page 19 INPUT POWER LINE SWITCH ELECTRONIC TIMER WALL SUPPORT X RAY HEAD IN OUT IN GROUND HAND CONNECTION JUMPERS H...

Страница 7: ...ators IEC 601 2 28 Particular requirements for the safety of X assemblies and X ray tube assemblies for medical diagnosis 21 CFR 1020 30 Diagnostic x ray systems and their major components 21 CFR 1020...

Страница 8: ...the wall support or into the timer Cables to be disconnected are those leaving the connection block towards the tube head and not those coming in from the timer Make sure that the disconnected cables...

Страница 9: ...type 93 200 01700 Anode voltage 70 kVp 8 at nominal line vo 66 kVp 8 at nominal line voltage 74 kVp 8 at nominal line voltage 10 Anode current 7 0 mA 15 at at nominal line voltage 5 3 mA 15 at nominal...

Страница 10: ...ght on hand switch and on control panel plus acoustic buzzer Preheating time Set at installation typical value 180 ms Hand switch Hand switch with 3 m coiled cord remote mounting possibility Overall s...

Страница 11: ...bile Stand 29 74 cm width 42 107 cm height 24 62 cm depth 65 165 cm total heath with folding arm 2 6 Weights Timer 3 7 lb 1 7 kg Tube head 14 1 lb 6 4 kg Round BLD 0 22 lb 0 1 kg Rectangular BLD 0 44...

Страница 12: ...wo bolts only top and bottom central holes T2 B2 are used when there is a solid slim column e g iron mounting with weak sides e g wooden wall Considering the requested safety factor each one of the tw...

Страница 13: ...25 ISO 8 8 M 8X1 ISO 8 8 5 16 18 UNC SAE Grade 5 5 16 24 UNF SAE Grade 5 INSUFFICIENT WALL OR HARDWARE STRENGHT MAY CAUSE THE WALL MOUNT TO PULL OUT FROM THE WALL AND THE FULL SYSTEM TO FALL ON TO TH...

Страница 14: ...nted far from the timer provided the is in view and the sound of the buzzer can be heard 3 1 4 Mounting and Connecting Sequence A Mount the Wall Support B Mount the Timer C Mount the Extension Arm D M...

Страница 15: ...in the wall according to the applicable type of mounting a Two bolts b Four bolts c Six bolts d With or without reinforcing plate D Slide the logo strip out from the plastic cover to access the screw...

Страница 16: ...ension Arm A Unpack the extension arm and check for completeness of parts B Cut but do not remove the string to be used later on to pull the cable from the Folding Simple Arm through the Extension Arm...

Страница 17: ...grease the mounting pin at base of Folding Arm C Secure the driving string to the end of the electrical cable of the Folding Arm D Pull the other end of the string to drive the electrical cable of th...

Страница 18: ...es from the Timer additional cable to the block and to the Folding Simple Arm Tube head as reported in figure B Connect the ground cable on the Wall Support to the blocking bolt on the extension arm S...

Страница 19: ...the measurement eventually the full line voltage should be read between line and ground test lamp lights c Make sure no voltage is read between neutral in and ground if not check line voltage distrib...

Страница 20: ...holder remote box has a hook to hold the hand plug below where to connect the coiled cable A cable 10 m long is provided with a plug on one side to connect to the socket on the timer and free wires on...

Страница 21: ...slot on the outer pin of the connecting socket and tighten the locking screw F Put back the cover plate and block it with the four retaining screws 3 1 14 Mounting the Beam Limiting Device A Mount th...

Страница 22: ...ee label c Set the pre heating time according to type of X ray insert tube head assembly the measured value is marked in front of the tube head G Put back the line voltage fuse s H Put back all plasti...

Страница 23: ...dip 2 dip 3 Setting Tube Head Type 110 or 220 VAC 115 or 230 VAC 120 or 240 VAC 125 or 250 VAC Set pre heating time as indicated on label tube head dip 5 dip 6 Setting Filament Pre 100 ms 120 ms 140 m...

Страница 24: ...OR WATCH DOG LED FUSE 4 T 200 mA 5 x 20 POWER SUPPLY 24 V LED FUSE 3 T 200 mA 5 x 20 OUTPUT LINE TO TUBEHEAD INPUT MAINS LINE NEUTRAL LIVE JUMPER FUSE 2 MAINS SWITCH HANDSWITCH FUSE 1 FUSE 2 T8A 6 3x3...

Страница 25: ...he existing local line voltage 3 2 2 Room Preparation Make sure that a wall socket is available close to the dental chair in reach of the mobile X ray equipment to be connected Such a socket has to be...

Страница 26: ...Stand The following sequence to be followed A Put together the base elements and tighten screw and washer B Mount the column timer plate externally to the long legs see picture and tighten it with the...

Страница 27: ...nt for live and neutral connections 3 2 8 Connecting the Timer For the connection of the output cable to the timer please refer to section 0 at page 18 3 2 9 Activation of Fuse 2 Units provided with s...

Страница 28: ...th the Manufacturer Any defect or non compliance must be reported promptly to the Manufacturer or to its Local Agent ALWAYS DISCONNECT THE SYSTEM FROM THE LINE VOLTAGE SUPPLY USING THE GENERAL LINE VO...

Страница 29: ...eck friction during rotation of Extension Arm and adjust relevant screw if needed D Verify that the technical label with identification data outside the plastic cover at bottom is in place and is read...

Страница 30: ...stand A Disconnect the tube housing assembly keeping the arm fully open to avoid the risk of injuries fig 1 B Remove the snap out caps of the first arm section the blocking screws of the pins then th...

Страница 31: ...IntraOs 70 Service Installation Manual Version 4 0 69 500 00210 Fig 4 Fig 1 Fig 2 Fig 5 Fig 8 BENT ARM IN SHADE INDICATES WEARING OUT OF BASIC PIN 31 52 Fig 9 Fig 6 Fig 7 Fig 3...

Страница 32: ...IntraOs 70 Service Installation M 32 52 69 500 00210 Fig 10 Fig 13 Fig 15 Fig 17 Fig 11 Service Installation Manual Version 4 0 Fig 12 Fig 14 Fig 16 Fig 18...

Страница 33: ...1 about 5 cm free thread left F Lubricate moving parts fig 17 and fig 18 and the three pins to block the metal covers before their insertion G Mount the metal covers provisionally with friction screws...

Страница 34: ...Beam Limiting Device A Verify that the collimator is properly mounted and in case lock it firmly closed B Verify that the internal cone is not damaged Replace the collimator if necessary C Verify that...

Страница 35: ...nce measure the dose for exposure of 0 2 s nominal irradiation of 10 pulses at 50 Hz 12 at 60 Hz C If dose at 0 2 s is 1 10 of dose at 2 s the value set for PHT D If dose at 0 2 s is lower than 1 10 o...

Страница 36: ...y loaded The kVp level can be measured with a non invasive kVp meter placed in front of the BLD following the instructions in the user manual of the instrument measure can be done with an exposure tim...

Страница 37: ...ed a precision resistor of 1000 Ohm between points A1 and A2 on the small on the back of the tube head by monitoring the voltage across it tube current with an oscilloscope and counting the number of...

Страница 38: ...ply line Using a voltage generator of not more than 6 V 50 60 Hz a current of 25 A is applied from 5 s to 10 s through the earth terminal to any metal part which can be in touch to the patient operato...

Страница 39: ...ngle Fault Condition SFC with S1 OFF S5 ON live and neutral conductors in normal condition S5 OFF swap of live and neutral conductors Please note that 1V equals to 1mA Make sure values of leakage curr...

Страница 40: ...ing Device FFD 21 cm Output Field 6 cm diameter C rectangular Beam Limiting Device FFD 21 cm Output Field 3 5x4 6 cm D Side cap tube head holder E Coupling with Concentric Pin for SN lower than TK0501...

Страница 41: ...tension Side D Arm Cover Tube Head Side E Retaining Clip With Screw F Cable 2 8 m With Concentric Connector for SN lower than FA0501 G Ring Spacer D 25 With Hole H Snap Out Cap set of 8 I Friction Scr...

Страница 42: ...0 90 cm 93 100 17400 110 cm List of Spare Parts Figure EA Item Description A Large Ring Spacer D35 B Snap out Cap D25 set of 8 C Friction Screw Brass Cap M6x10 set of 3 Friction Screw Brass Cap M8x10...

Страница 43: ...Installation Manual Version 4 0 69 500 00210 6 4 Wall Support 93 100 11000 List of Spare Parts Figure WS Item Description A Plastic Cover Wall Mount B Logo Strip Blue X Figure WS A B 43 52 Code 76 190...

Страница 44: ...t of Spare Parts Figure MB Item Description A Transport Handle set of 2 B Cable Holder With Strain Relief C Hook For Cable D Wheel set of 2 E Wheel With Brake set of 2 Figure MB A B C D E Service Inst...

Страница 45: ...trol Board With Triac E Keyboard Control Panel F AutoSet Membrane G Hand Switch Without Cable H Coiled Cable I Logo Strip IntraOs 70 J1 230V Fuse F1 F2 T5A 250V 5x20 set of 10 115V Fuse F1 F2 T8A 250V...

Страница 46: ...00 17200 Extension Arm 90 cm 93 100 17300 Extension Arm 110 cm 93 100 17400 Folding Arm 93 100 12010 Tube Head G 120 V 93 200 01300 Tube Head G 230 V 93 200 01700 20 cm Round BLD 91 300 00020 20 cm Re...

Страница 47: ...nt permanent connected to line Lower Incisor Inherent Fil Lower Canine Premolar g g g g Focal Spot Lower Molar Fragile Handle With Care Bite Wing Interproximal Fear of Humidity Digital Receptor Up Do...

Страница 48: ...00 0 80 0 64 0 50 0 40 0 32 0 25 0 20 Lower Molar 1 25 1 00 0 80 0 64 0 50 0 40 0 32 0 25 0 20 0 16 Lower Premolar 1 00 0 80 0 64 0 50 0 40 0 32 0 25 0 20 0 16 0 12 Lower Incisor 0 80 0 64 0 50 0 40...

Страница 49: ...ff and on again A 07 Exposure push button pressed at power on Red lamp Alarm flashing Exposure inhibited By acknowledgement on the panel A 08 Exposure stopped by the operator Red lamp Alarm flashing S...

Страница 50: ...IntraOs 70 Service Installation M 50 52 69 500 00210 Appendix E Identification Labels Service Installation Manual Version 4 0...

Страница 51: ...IntraOs 70 Service Installation Manual Version 4 0 69 500 00210 Appendix F Cooling Curves 51 52...

Страница 52: ...IntraOs 70 Dental X ray Equipment Service Installation Manual English Edition Version 4 0 6950000210...