To access the wiring terminal, open the enclosure by removing the four
screws with the provided 3/16” hex key. The meter must be powered by
5 volts DC. Wattage must not exceed 5 watts. 20-24 AWG (American
wire gauge) shielded cable is recommended for signal output
connections. The MS-6 wiring compartment is equipped with two
communications cable liquid-tight cable gland grommets for cable
diameters from .190 to .205 inches.
Note:
The blank grommet plug should be used when the communications
cable grommet is not required.
Refer to page 21, section 11 of the provided Operating Manual for further
details.
STEP 5 Wiring
STEP 6 Mounting
For the MS-6 Chemical Feed Flowmeter
to operate properly the pipe must be full, therefore it
must be installed in a vertical position with fluid flowing in an upward direction.
To reduce large particles and bubbles that can reduce signal quality, it is recommended that a
strainer is installed at the inlet of the meter as shown.
Installation Requirements:
1.
Mounting Direction can cause inaccuracies!
Meter must be installed in a vertical plane with fluid flowing in an upward
direction to ensure accuracy.
2. Vibration and heavy loads will damage the meter!
Wall, floor and ceiling mounts and supports must be carefully aligned with
the meter body and sturdy enough to support the plumbing and prevent
vibration. Never allow the meter to support the weight of related piping.
3.
High pressures and temperatures will damage the meter!
The maximum acceptable temperature and pressure is interdependent. The
maximum acceptable working pressure is dependant on the actual fluid
temperature. The maximum acceptable fluid temperature is dependant on the
actual working pressure. For more information refer to the provided Operating
Manual (see Temperature Vs. Pressure chart on page 3).
4.
Strainer Recommended!
1/32” mesh strainer is recommended to be installed at the inlet of the meter.
This will keep particles from entering the meter that could cause inaccurate
readings.
3
3
1
2
2
2
Ceiling
1
Floor
Product Dimensions:
P.N. 80000-603 rev. 20190722
FLOW DIRECTION
STRAINER
3.7in
[9.4cm]
3.8in
[9.7cm]
3.9in
[9.9cm]
10.02in
[25.45cm]
3.4in
[8.6cm]
Quickstart Guide
MS-6 Chemical Feed Flowmeter