Pump
1. Wash hands and place gloves on hand.
2. Place pump, hose on decontamination table or counter.
3. Spray with disinfectant.  Do not use phenol based disinfectants.
4. Wipe down pump and hose 
5. Let pump and hose dry for approximately 15 minutes.

Mattress
1. Wash hands and use gloves.
2. Place mattress on decontamination table or counter.
3. Spray with a Quaternary disinfectant.  Do not use phenol based disinfectants.
4. Wipe down top of mattress with towel and let sit for approximately 15 minutes.
5. Turn mattress over and repeat the cleaning process.
6. Let the mattress hang dry in a well-ventilated area between 12 - 24 hours.
7. Covers can be removed and washed in a gentle cycle using disinfectants. 
8. Put in dryer only in spin cycle to remove excess moisture, do not use heat.

MAINTENANCE

TROUBLESHOOTING

PROBLEM

POSSIBLE CAUSES

ACTION

Mattress Not Inflating

1. No Power

1. Check plug in outlet, check fuse

2. CPR Valve opened

2. Twist to close position

3. Air Rail or Hoses Kinked

3. Check that hoses are not impeded

4. Hose not connected to pump

by bed rail or other obstruction

Power Failure

1. Power Surge

1. Turn power switch off then on 

2. Blown fuse

2. Change fuse

3. Power cord is not plugged in

3. Plug power cord into outlet

Pump Not Inflating

1. Pump in standby mode

1. Press operate button, select 

auto firm, or alternate mode.

2. Not responding to mode

2. Turn Power switch on then off

to reset.

3. Filter is dirty

3. Remove filter located under the

carry handle of pump. Vacuum
or rinse under water to remove 
excess dust.

4. Excess Noise

4. Return for authorized technical

service.

Medical Products Inc.

Blue Chip

7-11 Suffern Place, Suffern, NY  10901

Toll Free: 800 795-6115  •   Fax: 845 369-7633

Website: www.bluechipmedical.com

Pump
1. Wash hands and place gloves on hand.
2. Place pump, hose on decontamination table or counter.
3. Spray with disinfectant.  Do not use phenol based disinfectants.
4. Wipe down pump and hose 
5. Let pump and hose dry for approximately 15 minutes.

Mattress
1. Wash hands and place gloves on hand.
2. Place mattress on decontamination table or counter.
3. Spray with Madacide or other disinfectants.  Do not use phenol based disinfectants.
4. Wipe down top of mattress with paper towel and let sit for approximately 15 minutes.
5. Turn mattress over and repeat the cleaning process.
6. Let the mattress hang dry in a well-ventilated area between 12 - 24 hours.
7. Covers can be removed and washed in a gentle cycle using disinfectants. 
8. Put in dryer only in spin cycle to remove excess moisture, do not use heat.

MAINTENANCE

TROUBLESHOOTING

PROBLEM

POSSIBLE CAUSES

ACTION

Mattress Not Inflating

1. No Power

1. Check plug in outlet, check fuse

2. CPR Valve opened

2. Twist to close position

3. Air Rail or Hoses Kinked

3. Check that hoses are not impeded

4. Hose not connected to pump

by bed rail or other obstruction

Power Failure

1. Power Surge

1. Turn power switch off then on 

2. Blown fuse

2. Change fuse

3. Power cord is not plugged in

3. Plug power cord into outlet

Pump Not Inflating

1. Pump in standby mode

1. Press operate button, select 

auto firm, or alternate mode.

2. Not responding to mode

2. Turn Power switch on then off

to reset.

3. Filter is dirty

3. Remove filter located under the

carry handle of pump. Vacuum
or rinse under water to remove 
excess dust.

4. Excess Noise

4. Return for authorized technical

service.