Datum 'L' Vaporizer
User Manual
24
WARNING: Tighten the drain plug securely before replacing
the filler cap.
WARNING: Anaesthetic agent drained from the vaporizer
must be discarded and not reused. Treat as a hazardous
chemical.
4.2 Keyed Filler
4.2.1 Filling
1. Ensure the filler control is turned fully counterclockwise
(closed).
2. Loosen the clamp screw and remove the slipper block from
the filler port.
3. Screw the filler adaptor on to the supply bottle and tighten it
fully to ensure it is airtight.
4. Insert the end of the filler adaptor until it stops, into the filler
port on the vaporizer. If an incorrect adaptor has been fitted
to the supply bottle it will not be possible to insert it into the
port.
5. Tighten the clamp screw and ensure the adaptor is secured.
6. Raise the bottle to a level above the filler port so that any air
in the bottle cannot enter the filler tube.
7. Open the filler control port by pulling the lever clockwise until
it stops.
8. Fill the chamber to the required level shown on the indicator.
WARNING: DO NOT OVERFILL. If the vaporizer is overfilled it
must be withdrawn from service.
WARNING: DO NOT LOOSEN THE FILLER ADAPTOR FROM
THE DRUG BOTTLE DURING FILLING AS OVERFILLING MAY
OCCUR.
WARNING: Do not tip the vaporizer during the filling
operation as overfilling may occur.
Содержание Datum L Series
Страница 1: ...User Manual Datum L Vaporizer 1 Blease User Manual Blease Anaesthesia Vaporizer R L Series ...
Страница 2: ...Datum L Vaporizer User Manual 2 ...
Страница 4: ...Datum L Vaporizer User Manual 4 Read this Manual before operating the ventilator ...
Страница 13: ...User Manual Datum L Vaporizer 13 Symbols and abbreviations used on Blease Equipment ...
Страница 14: ...Datum L Vaporizer User Manual 14 Notes ...
Страница 35: ...User Manual Datum L Vaporizer 35 Notes ...