Static Alignment
Fully extend limb to access the adjuster.
Rotate the SPRING STACK ADJUSTER to achieve
the required stability for the individual amputee.
Note: Audible clicks should be heard upon
adjustment, and the adjuster is self locking.
Tighten the SPRING STACK to reduce stability.
Release the SPRING STACK to increase stability -
See above.
If required, the adjuster may be reset to the
factory setting. This can be achieved by reducing
stability by six audible clicks from the maximum
stability setting.
Build Line
The instructions in this section are for practitioner use only.
1.
Complete static alignment as per
diagram below.
ESK+ function must be confirmed by
wearer before walking on the limb.
(See
Alignment Advice and Stability
Spring Stack Adjuster
Underside View of Chassis
Decrease
Stability
Increase
Stability
Trochanter
Knee Pivot
Ankle Centre
2.
Walking Trials.