Select the Units of Measure
The monitor is programmed with five common rate/total unit options The monitor shows the rate/total unit that the display
is currently set for If the current selection is correct, press
ENTEO
once to advance to the
DISPLAY FUNCTION
parameter
To change the unit, press
UP
or
OIGHT
to scroll to the correct rate unit Press
ENTEO
to save and advance to the
DISPLAY
FUNCTION
parameter
Selection
Oate
Total
GPM/GAL
Gallons per minute
Gallons
LPM/LIT
Liters per minute
Liters
M3PH/M3
Cubic meters per hour
Cubic meters
M3PD/M3
Cubic meters per day
Cubic meters
BPD/BBL
Oil barrels per day
Oil barrels
Select the Display Function
The monitor displays
RATE
,
TOTAL
, alternates between
BOTH
rate and total or
TEST
If the current selection is correct, press
ENTEO
to advance to the next parameter To change to an alternate display mode, press
UP
or
OIGHT
to scroll to the correct
display mode and press
ENTEO
to save and advance to the
TOTALIZER PULSE OUTPUT
parameter
The
TEST
function acts like a frequency counter and displays the raw input frequency being supplied to the frequency input
terminals This is very useful when troubleshooting flow problems
Totalizer Pulse Output
The pulse output parameter can be enabled or disabled When enabled this output generates 20 mS duration pulse for every
time the least significant digit of the totalizer increments The amplitude of the pulse is dependent on the voltage level of the
supply connected to the pulse output and is limited to a maximum 30V DC Press
UP
or
OIGHT
to scroll to
ON
or
OFF
Press
ENTEO
to save and advance to the
PASSWORD
parameter
Password
The
PASSWORD
setting restricts access to
PROGRAM
mode Initially, the password is set to all zeros and any user can modify
the parameter settings in
PROGRAM
mode To change the password, enter any four digit code Press
UP
to increment the digit,
and
OIGHT
to advance to the next digit Press
ENTEO
to store the password and advance to
RST PSWD
The new password is
now required to enter
PROGRAM
mode With this password set, any user is able to reset the stored totals on the monitor
Oeset Password
The
RST PSWD
restricts resetting the totals on the monitor The
PASSWORD
must also be set to restrict the total reset Initially,
the password is set to all zeros and any user can reset the stored totals on the monitor To change the password, enter any four
digit code Press
UP
to increment the digit, and
OIGHT
to advance to the next digit After entering all digits, press
ENTEO
to
store the password and return to
RUN
mode The reset password is now required to reset the totals on the monitor
Oun Mode
Oeset Total
To reset the monitor total display, in
RUN
mode press
MENU
and
ENTEO
simultaneously If the
RST PSWD
has been set to
something other than the default, you must enter the new password before the monitor will reset The monitor flashes
TOTAL RST
and returns to
RUN
mode at the conclusion of the reset procedure
Store Total
The current total can be manually stored in the monitor’s flash memory Use this feature prior to replacing the battery Press
and hold
ENTEO
for 2 seconds The display responds with a flashing
TOTALSVD
and then returns to
RUN
mode
Automatic Store Total
The monitor is equipped with a store total feature that works automatically, saving the current total to flash memory The
frequency of saves depends on the power supply option chosen
Battery Powered: Once per hour and just before a low battery condition turns the unit off
Loop Powered: Once every ten minutes
Programming
Page 7
March 2017