Revision 7.00 / LB-0462 – CerePlex Direct Instructions for Use 
 

 

15

Disposal 

All devices, both used and unused, should not be disposed with household 

waste. Return to a recycling point for electric and electronic devices. 

Magnetic Resonance 

The CerePlex Direct has not been evaluated for safety and compatibility in the MR environment. 

The CerePlex Direct has not been tested for heating, migration, or image artifact in the MR 

environment.  

Troubleshooting 

No CerePlex Direct Detected 

 

Figure 9 - Error Message Showing No CerePlex Direct Detected 

 

This error indicates an error has occurred in communication between nPlayServer and 

Central Software Suite. If the CerePlex Direct and the PC are connected through the 

USB cable and the CerePlex Direct is powered on, this may be caused by an error with 

the USB driver.  
To correct this, uninstall the Total Phase USB driver in the Control Panel’s Programs 

and Features menu.

 

After the uninstall is complete, run CDM21224_Setup.exe from the installation directory 

of the software. 

 

Содержание CerePlex Direct

Страница 1: ...uite 200 Salt Lake City UT 84108 USA P 1 801 582 5533 F 1 801 582 1509 www blackrockmicro com Revision 7 00 LB 0462 CerePlex Direct Instructions for Use 2019 08 2019 Blackrock Microsystems LLC CerePle...

Страница 2: ...are 10 CerePlex Direct Inputs 11 Digital Headstage Input Banks 11 Analog Input 12 Digital Input 12 Fast Settle Input 12 CerePlex Direct Outputs 12 USB Data Line to PC 12 Patient Ground 13 NeuroMotive...

Страница 3: ...CerePlex Direct Instructions for Use 3 Digital Headstage Connectivity 16 Return Merchandise Authorizations 17 Warranty 17 Support 17 Manuals Software Downloads and Application Notes 18 Complaints 18 I...

Страница 4: ...e of recording from 16 32 64 96 128 or 256 channels over 3 headstage input banks depending on the model Figure 1 CerePlex Direct Data Acquisition System This manual will cover the system requirements...

Страница 5: ...iquids Contact with water shower spray or wet surfaces can result in a device malfunction and or electrical shock Connection of external instruments to the CerePlex Direct may compromise electrical sa...

Страница 6: ...Model Name CerePlex Direct Digital Signal Processing Adaptive line noise cancellation filter 4th order hi lo pass digital filtering for all channels Separate filters for simultaneous continuous and sp...

Страница 7: ...ed headstages Common Mode Rejection Ratio 90 dB at 50 60 Hz Experimental Digital Inputs Sixteen 0 5 V TTL Logic Level inputs for experiment event signaling Power Supply Standard 3 pin PC power connect...

Страница 8: ...s the manufacturer s serial number so that a specific medical device can be identified 5 2 8 Do Not Use if Package is Damaged Indicates that a medical device should not be used if the package has been...

Страница 9: ...Software Suite with the CerePlex Direct connected to the host PC may result in driver installation issues and cause the host PC to fail to recognize the CerePlex Direct If this happens please refer to...

Страница 10: ...a bandwidth of 50 Mbps The CerePlex Direct is designed to have several isolated ground planes The CerePlex Direct chassis is tied to earth ground through the ground pin of the power plug Patient grou...

Страница 11: ...ed A B and C for digital headstage inputs There is an LED on each of the digital headstage input banks that will be lit blue when the Direct is receiving valid data packets from the digital headstage...

Страница 12: ...to Pin 14 on the digital data connector so that downstream digital headstages can receive a fast settle signal The two other BNC connectors are labeled NC and reserved for future feature use CerePlex...

Страница 13: ...nce connection to patient ground NeuroMotive Camera Sync If using the NeuroMotive video tracking system the CerePlex Direct can be hardware synchronized with the tracking and recording system The BNC...

Страница 14: ...terface is identical to Central For more detailed instructions and use on Central Software Suite please see LB 0574 Central Software Suite IFU Cleaning Maintenance and Disposal Cleaning The CerePlex D...

Страница 15: ...act in the MR environment Troubleshooting No CerePlex Direct Detected Figure 9 Error Message Showing No CerePlex Direct Detected This error indicates an error has occurred in communication between nPl...

Страница 16: ...Error Message In rare cases USB hardware can have compatibility conflicts with the CerePlex Direct causing periodic packet loss between the Direct and host PC If this happens try switching to a differ...

Страница 17: ...Microsystems Blackrock warrants its products are free from defects in materials and manufacturing for a period of one year from the date of shipment At its option Blackrock will repair or replace any...

Страница 18: ...pport Complaints When filing a complaint please provide the product description product number software version lot number complainant s name and address and the nature of the complaint Issues or Ques...

Отзывы: