488 CONTROLLER
12
3.1 Configuration
Three internal DIP switches set the interface’s configuration.
NOTE
Selectable functions are read
only
at power-on and should only be set prior to applying
power to the interface.
The figures below and on the next page illustrate the factory-default conditions, which are:
Serial Port:
IEEE:
9600 Baud
Mode = System Controller
8 Data Bits
Address = 10
2 Stop Bits
Bus Terminator = CR-LF; EOI Disabled
No Parity
Talk-Back Enabled
Serial Terminator = CR-LF
Echo Disabled
RTS/CTS Handshake
Figure 3-1. SW3 factory-default settings.
1 2 3 4 5 6 7 8
OPEN
SW3
IEEE Addr
IEEE Term
EOI
10
CR-LF
Disabled
Switch
Side
View
DOT
3. Getting Started