45
10 programs that can have 150 steps in total for setting purpose. This can only
be programmed remotely via USB, GPIB, or LAN with SCPI commands or
with the included software. Each program does not restrict the step quantity,
however, the sum of 10 programs are restricted to 150 steps. For each program
users can set up how many times to repeat the program. For each step users
may be able to set up the output voltage, current, and period of time (50 ms
minimum) to stay on the step. Please refer to “SCPI command subsystem”
section for detail SCPI commands.
Below are some examples of commands used to setup a custom program in list
mode.
Example 1:
To output the wave form shown above, users may edit the program as the
following orders:
PROG 1
Choose program number
PROG:CLE
Clear program 1 data
PROG:REP 0
No repeat (repeat one time for “1”)
PROG:TOTA 8
Set program 1 to have 8 steps in total
PROG:STEP 1
Following 3 settings are for step 1
PROG:STEP:CURR 1
Set output current to 1 ampere
PROG:STEP:VOLT 5
Output voltage is set to 5 volts
PROG:STEP:ONT 0.1
Output ON time is set to 0.1 sec
PROG:STEP 2
Following 3 settings are for step 2
PROG:STEP:CURR 1
PROG:STEP:VOLT 10