55
12.
Press Esc key to exit.
CH01 CC OFF
Vdc = 0.0000V
Adc = 0.0000A
Wdc = 0.00W
13.
Press Shift + ⑦ to enter System menu. Press
key to select <Trigger
source> and press Enter. Select <Manual> and press Enter to confirm.
MANUAL
EXTERNAL
HOLD
BUS
TIMER
14.
Press Esc key to exit.
CH01 CC OFF
Vdc = 0.0000V
Adc = 0.0000A
Wdc = 0.00W
15.
Press On/Off key to turn on the electronic load’s input.
CH01 CC ON
Vdc = 0.0000V
Adc = 0.0000A
Wdc = 0.00W
16.
Press Tran key on channel 1 module to enable transient operation.
17.
Press Trig key on mainframe panel to trigger transient operation.
Transient Operation Programming Example
If in remote control mode, you can use the following example commands for transient operation
(refer to MDL Series Programming Guide for more information).
CURRent:TRANsient:MODE CONTinuous
CURRent:TRANsient:ALEVel 10
CURRent:TRANsient:AWIDth 0.2ms
CURRent:TRANsient:BLEVel 2
CURRent:TRANsient:BWIDth 0.3ms
TRANsient ON
TRIGger:IMMediate