CHAPTER 2. SETUP
12
Figure 2.10: Offset Calibration Screen
2.6 Calibration
You can easily calibrate the recorder inputs for the voltage and thermocouple offsets. This will remove any offset
present on different channels.
To do so:
1. Let the unit work for 20 minutes (ambient temperature 68-77F (20-25C))
2. Enter the recorder main page and then press the “Setup” key
3. Select the “Additional Option” menu
4. Select the “Electric Calibration” menu
5. Press the “Calibration offset” to start calibration. This takes approximatly 5 minutes
6. Restart the unit
2.6.1 Factory Default Calibration
You can restore the default factory calibration to correct any possible mistake in the calibration coefficients.
This restores calibration values configured at the factory. This is performed as follows:
1. Let the unit work for 20 minutes (ambient temperature 68-77F (20-25C))
2. Enter the recorder main page and then press the “Setup” key
3. Select the “Additional Option” menu
4. Select the “Electric Calibration” menu
5. Press the “Recover factory calibration”
6. Restart the unit.