Sound Level Calibration 

 

 

6. Operating Preparation 

(1) Remove battery cover and install a 9V Battery in 

the battery compartment. 

(2) To quickly check the operation of the sound level 

calibrator.  

Proceed as follows: 
(a) Turn the power switch from OFF to the 94dB 

position. The user can start to operate it when 
LED indicates a green light. If there is no light, 
please replace the battery. 
(b)Change the switch from 94dB to 114dB, the dB 

level will increase by 20. You can calibrate in noisy 

environments and check linearity.   

7. Calibration Procedure   

 

 

 

(1) The cavity of the calibrator will accommodate 1-inch 

microphone. 

(2)  When the calibration is performed to an 

instrument with 1/2-inch microphone, the 1/2-inch 
microphone adaptor will have to be inserted by 
gently pushing it into the cavity till the end. 

(3) Place the sound level calibrator over the 

microphone of the sound measuring instrument 

ANSI S1.40 - 1984

IEC 60942 Class 2

114dB

94dB

OFF

Lo  =

30

80 dB

Med= Hi   =

POWER-UP  OPTIONS

Auto=

100 dB 130 dB

130 dB

50 80

30

MAX

CLOCK

REC

MIN

LEVEL

SLOW

INTV

FAST

SETUP

C

A

www.

.com

[email protected]

1.800.561.8187

Содержание CAL73

Страница 1: ...tput frequency 2 Reference conditions 2 Influence of ambient conditions 2 Total Harmonic Distortion THD 2 Accuracy of sound pressure level 2 Power 2 Battery life 2 Battery test 3 Dimensions 3 Weight 3 Ambient conditions 3 Storage temperature humidity 3 Accessories 3 5 Nomenclature And Function 4 6 Operating Preparation 5 7 Calibration Procedure 5 www com information itm com 1 800 561 8187 ...

Страница 2: ...ed in this manual should only be performed by qualified personnel Periodically wipe the case with a dry cloth Do not use abrasives or solvents on this instrument Safety symbols Comply with EMC When servicing use only specified replacement parts 2 Description Sound level calibrator is used to calibrate sound level meters and other sound measurement equipment You can calibrate 1 inch diameter microp...

Страница 3: ...ncy 1000Hz 2 Reference conditions Temperature 23 C 74 F Relative humidity 50 Atmospheric pressure 1013hpa Influence of ambient conditions Temperature coefficient 0 005dB Humidity coefficient 0 005dB RH Total Harmonic Distortion THD 3 Accuracy of sound pressure level 0 5dB Power one 9V battery 006P or IEC 6F22 or NEDA 1604 Battery life approx 40 hours Alkaline Battery Battery test Internal circuitr...

Страница 4: ...e acceptable range Dimensions LxWxH 4 4 x 2 4 x 1 7inch 113 x 63 x 44mm Weight approx 170g inc luding battery Ambient conditions 0 40 C 32 140 F 10 90 RH 650 1080hpa Storage temperature humidity 10 50 C 14 122 F 0 70 RH Accessories Instruction manual carrying case 9V battery 1 2 microphone adapter www com information itm com 1 800 561 8187 ...

Страница 5: ...cate LED Power and output level select switch ƒ Transducer assembly 1 inch cavity for microphone insertion 1 2 inch microphone adapter Battery cover 94dB 114dB ANSI S1 40 1984 IEC 60942 Class 2 OFF NEDA 1604 9V BAT TERY 6F22 006P 1 2 3 4 5 www com information itm com 1 800 561 8187 ...

Страница 6: ...20 You can calibrate in noisy environments and check linearity 7 Calibration Procedure 1 The cavity of the calibrator will accommodate 1 inch microphone 2 When the calibration is performed to an instrument with 1 2 inch microphone the 1 2 inch microphone adaptor will have to be inserted by gently pushing it into the cavity till the end 3 Place the sound level calibrator over the microphone of the ...

Страница 7: ...rument may be set to FAST or SLOW response and C or A weighting 5 Read the level on the instrument under test and adjust the sensitivity control for the correct indication of the sound level calibrator level selected in step 6 When the calibrator is not being used please switch OFF the power to save the battery CAUTION Ambient sources of noise or vibration can cause a false calibration indication ...

Отзывы: