Example: A power supply is connected to the DC load with 72.5 cm of 20-gauge solid copper wires.
The constant current i set to 5 A. The power supply's output meter reads 27.0 V and the DC Load's
voltage display reads 26.71 V with a power dissipation of 133.70 W. This is without remote sensing
enabled. With remote sensing turned on and the remote sensing terminals connected to the power
supply output terminals, the DC Load reads 26.98 V and indicates 134.95 W of power dissipation.
This demonstrates that (134.95 - 133.70) = 1.25 W is dissipated in the 20-gauge wires. From the
resistance per unit length of 20-gauge wires, the expected dissipated power at 5 A can be calculated
as 1.21 W.
Saving and Recalling settings
The DC Load provides 25 non-volatile registers to save instrument settings for recall later.
To save the instrument's settings to a register, press
Shift
+
Store
. You'll be prompted for a register
number. Enter a number between 1 and 25, then press the
Enter
key. The settings are saved.
Note that this will overwrite any values previously saved in that register.
To recall the instrument's settings from a register, press
Shift
+
Recall
. You'll be prompted for a
register number. Enter a number between 1 and 25, then press the
Enter
key. The stored settings
are saved. Note that this will overwrite all values currently set in the instrument.
If you try to recall a register with no data in it, you will get a
NO EEPROM DATA
error message.
8500 DC Load Series
Version: February 4, 2009
Page 43 of 76
Содержание 8500
Страница 1: ...USER MANUAL 8500 SERIES DC ELECTRONIC LOADS Models 8500 8502 8510 8512 8514 8518 8520 8522 8524 8526 ...
Страница 24: ...Models 8510 8512 8514 8518 Models 8520 8522 8500 DC Load Series Version February 4 2009 Page 24 of 76 ...
Страница 25: ...Models 8524 8526 Unit mm 8500 DC Load Series Version February 4 2009 Page 25 of 76 ...
Страница 70: ...Byte offset Meaning 12 24 Reserved 8500 DC Load Series Version February 4 2009 Page 70 of 76 ...