![BK Precision 817050088922 Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/bk-precision/817050088922/817050088922_manual_2759787005.webp)
6
2. AC Voltage Measurement (True RMS):
Range
: 2V, 20V, 200V, and 600V
Resolution
: 1mV, 10mV, 100mV, and 1V
Accuracy
: +/- (0.8% + 5 digits)
Input Impedance
: 1 M-Ohm
3. DC Current Measurement:
Range
: 2mA, 20mA, 200mA, and 2000mA
Resolution
: 1uA, 10uA, 100uA, and 1mA
Accuracy
: +/- (0.4% + 3 digits)
Current Shunt
: 0.1 Ohm @ >20mA, 10 Ohm @
20mA
4. AC Current Measurement (True RMS):
Range
: 2mA, 20mA, 200mA, and 2000mA
Resolution
: 1uA, 10uA, 100uA, and 1mA
Accuracy
: +/- (0.8% + 5 digits)
Current Shunt
: 0.1 Ohm @ >20mA, 10 Ohm @
20mA
Note:
The accuracy of DC/AC voltage/current measurements is only applied when in 5% - 100% of the range.
5. LCR Measurement:
Z Accuracy (
Ae
):
|Zx|
Freq.
20M ~
10M
(
)
10M ~
1M
(
)
1M ~
100K
(
)
100K ~
10K
(
)
10K ~
1K
(
)
1K ~ 100
(
)
100 ~ 1
(
)
1 ~ 0.1
(
)
DCR
2% ±1
1% ±1 0.5% ±1 0.2% ±1 0.1% ±1 0.2% ±1 0.5% ±1 1% ±1
100Hz
120Hz
1KHz
10KHz
5% ±1
2% ±1
100KHz
200KHz
NA
5% ±1
2% ±1
1% ±1 0.4% ±1 1% ±1 2% ±1 5% ±1
Note:
1. The accuracy applies when the test level is set to 1Vrms.
2
. Ae
multiplies 1.25 when the test level is set to 250mVrms.
3
. Ae
multiplies 1.50
when the test level is set to 50mVrms.
4. When measuring L and C, multiply
Ae
by
2
1
Dx
if the Dx
>
0.1.
:
Ae
is applied only when the test level is set to 1Vrms.
www.
GlobalTestSupply
.com
Find Quality Products Online at: