32
CB command
Syntax:
CBx [y]
Parameters:
x represents a digit that has only 2 valid values: “I” or “S”.
y represents a decimal digit, valid if x = S.
Description:
The command sets/interrogates the mode of displaying the decimals in American or
French.
If x = S and y = 0, the American mode of displaying the decimals using point will be
engaged.
If x = S and y = 1, the French mode of displaying the decimals using comma will be
engaged.
Example:
CBS0 – the decimals will be displayed according to the American mode.
Message returned to terminal:
“
LCD point display engaged (American option
)” if y = 0.
“LCD comma display engaged (French option)”
if y = 1.
“Incorrect command”
if any of the digits sent was not valid.
CG command
Syntax:
CGx[y]
Parameters:
x represents a digit that has only 2 valid values: “I” or “S”.
y represents a digit that has 2 valid values: “C” or “T”, only if x = “S”.
Description:
The command sets/interrogates the output type on
TTL/CMOS
jack.
If x = “I”, the type of signal outputted on
TTL/CMOS
jack will be send to the
terminal.
If x = “S” and y = “C”, a
CMOS
style signal will be outputted on
TTL/CMOS
jack.
If x = “S” and y = “T”, a
TTL
style signal will be outputted on
TTL/CMOS
jack
Example:
CGST
Message returned to terminal:
“
TTL signal on TTL/CMOS jack” if y = “T”
“CMOS signal on TTL/CMOS jack” if y =”C”
“Incorrect command” if any of the digits sent was not valid.
CA command
Syntax:
CAx[y]
Parameters:
x represents a digit that has only 2 valid values: “I” or “S”.
y represents a digit that has 2 valid values: “E” or “D”, only if x = “S”.
Description:
The command sets/interrogates if –20dB attenuation is engaged or disengaged on the
main signal output,
OUTPUT
jack.
If x = “I”, a message indicating if attenuation is engaged or disengaged at signal
output will be send to the terminal.
If x = “S” and y = “E”, the – 20dB attenuation will be engaged on
OUTPUT
jack.
If x = “S” and y = “D”, the – 20dB attenuation will be disengaged on
OUTPUT
jack.
Example:
CASE – will engage the – 20dB attenuation.