7. A superimposed DC component can be added to the output signal 

by engaging the 

DC OFFSET

 switch (12) to enable operation of 

the 

DC OFFSET

 control (5). Rotation of this control adds a 

positive or negative DC component to the output signal. The DC 
component introduced is independent of the 

OUTPUT LEVEL

 

control and can be varied by ±10 volts open circuited or ±5 volts 
into 50

. The DC Offset does not affect the 

TTL/CMOS

 output 

jack. The effect of DC OFFSET is shown in Fig. 3. 

CONSIDERATIONS 

1.

 

l. Counterclockwise rotation of the 

COARSE

 frequency control 

decreases the output frequency to approximately one-tenth of the 
maximum for the range selected (10:1). For example, if the 

50K

 

range is selected and the 

COARSE

 frequency control is set to 

full counterclockwise, the output frequency is approximately 

5kHz

2.

 

It is advisable to set the 

FINE

 frequency control to the 

approximate center of its rotation before setting the 

COARSE

 

frequency control. This assures that the 

FINE

 control will not 

reach its limit while trying to finalize the frequency setting. 

3.

 

The 

FINE

 frequency control provides approximately ±5% 

frequency deviation from the 

COARSE

 control setting. This 

provides vernier adjustment to easily set the frequency to a 
precise value. 

4.

 

. When the 

5Hz

 range is selected, the gate time is 10 seconds and 

the display is updated once every 10 seconds. The result of a 
frequency change will not be displayed until 10 seconds later. 
Adjust the frequency in progressively smaller steps, waiting for 
the display to update until the desired frequency is obtained. 

5.

 

When outputting square waves or when using the TTL output, 
terminate the cable into 50

 to minimize ringing. Also, keep 

cables as short as possible. 

 

Figure 3. Use of DC OFFSET Control 

6. Remember that the output signal swing of the 
generator is limited to ±10 volts open circuited or ±5 
volts into 50

, and applies to the combined peak-to-

peak signal and DC offset. Clipping occurs slightly 
above these levels. Fig. 3 illustrates the various 
operating conditions encountered when using the DC 
offset. If the desired output signal is large or if a large 
DC offset is required, an oscilloscope should be used 
to make sure that the desired signal is obtained without 
undesirable clipping.

 

Содержание 4011A

Страница 1: ...1 INSTRUCTION MANUAL MANUAL DE INSTRUCCIONES MODEL 4011A MODELO 4011A 5 MHz FUNCTION GENERATOR with DIGITAL DISPLAY 5MHz GENERADOR DE FUNCIONES CON DIGITOS INDICADOR...

Страница 2: ...pose high voltage needlessly Remove housings and covers only when necessary Turn off equipment while making test connections in high voltage circuits Discharge high voltage capacitors after removing p...

Страница 3: ...3 Instruction Manual for MODEL 4011A 5 MHz FUNCTION GENERATOR with DIGITAL DISPLAY 22820 Savi Ranch Parkway Yorba Linda CA 92887 4610...

Страница 4: ...ONS 10 Frequency and Waveform Selection 10 Considerations 11 Duty Cycle Control 12 TTL CMOS Output 13 Voltage Controlled Frequency Operation 13 Output Protection Guidebook 14 page MAINTENANCE 15 Fuse...

Страница 5: ...e engineering manufacturing servicing educational and hobbyist fields The heart of the function generator is a VCG voltage controlled generator that produces precision sine square or triangle waves ov...

Страница 6: ...led Generator Frequency Stability Output frequency will not change more than 0 09 in 15 minutes after 1 hour warmup OUTPUT CHARACTERISTICS Impedance 50 10 Level 20V p p Open circuit 10V p p into 50 Am...

Страница 7: ...p Continuously variable Rise Time 120nS VCG Voltage Controlled Generator INPUT Input Voltage 0 10V 1V causes a 100 1 frequency change Impedance 10 k 5 FREQUENCY COUNTER Accuracy Time Base Accuracy 1 c...

Страница 8: ...ion from center changes the DC offset in a negative direction 6 OUTPUT Jack Waveform selected by FUNCTION switch as well as the superimposed DC OFFSET voltage is available at this jack 7 TTL CMOS Jack...

Страница 9: ...t frequency from 0 1 to 1 times the selected range 17 COUNTER DISPLAY Displays frequency of internally generated waveform 18 GATE LED Indicates when the frequency counter display is updated When the 5...

Страница 10: ...f the FUNCTION switches 3 Phase relationships of the waveforms are shown in Fig 2 4 Select the frequency of the waveform by engaging one of the RANGE switches 2 The output frequency is displayed along...

Страница 11: ...while trying to finalize the frequency setting 3 The FINE frequency control provides approximately 5 frequency deviation from the COARSE control setting This provides vernier adjustment to easily set...

Страница 12: ...produced The Model 4011A provides for symmetry variation from 15 to 85 1 Select the waveform desired either SINE SQUARE or TRIANGLE 2 Engage the DUTY CYCLE switch 14 and adjust the DUTY CYCLE control...

Страница 13: ...tation For example if the starting frequency is 500kHz applying 10V will change the output frequency to 5kHz 3 To operate the function generator as a sweep generator apply a positive going ramp signal...

Страница 14: ...rcuit under test and adjust the CMOS LEVEL control as instructed in the manual 5 When the function generator is used by students or other inexperienced users the circuit of Fig 5 could be added into y...

Страница 15: ...cedures FUSE REPLACEMENT 1 Locate the fuse holder on the input line receptacle 2 Remove the fuse holder and replace the fuse with an equal value fuse INSTRUMENT REPAIR SERVICE Because of the specializ...

Страница 16: ...nt Technical advice on selecting the best instrument for a given task Information on optional accessories for your instrument Information on instrument repair and recalibration services Replacement pa...

Страница 17: ...ccount must include payment in the form of a money order or credit card For the most current repair charges please visit www bkprecision com and click on service repair Return all merchandise to B K P...

Страница 18: ...sion com within fifteen 15 days of purchase Exclusions This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized alterations or repairs The warranty is...

Страница 19: ...f most test instruments to a hot chassis To test hot chassis equipment always connect an isolation transformer between the ac outlet and the equipment under test The B K Precision Model TR 110 or 1604...

Страница 20: ...nexiones en circuitos de alto voltaje Descargue los capacitadotes de alto voltaje despu s de remover poder 2 Si es posible familiar cese usted mismo con el equipo que va ha ser revisado y los lugares...

Страница 21: ...21 MANUAL DE INSTRUIONES MODELO 4011A 5MHz GENERADOR DE FUNCIONES CON DIGITOS INDICADOR 22820 Savi Ranch Parkway Yorba Linda CA 92887 4610...

Страница 22: ...deraciones 29 Control de ciclo de servicio 30 Salida para TTL CMOS 31 Operaci n de voltaje controlada frecuencia 31 Consideraciones de protecci n de salida 31 Gu a de aplicaciones del generador de fun...

Страница 23: ...el generador de funciones es un VCG voltaje controlado generador que produce ondas de seno cuadradas y triangulares desde los 0 5Hz hasta 5MHz rango Esto en compasa subaudible sonido untrasonido ultr...

Страница 24: ...ido Estabilidad de frecuencia la salida combiara menos que 0 09 sobr 15 minutes despu s de 1 hora de caletamlento CARACTER STICAS DE SALIDA Impederica 50 OHM 10 Nivel 20Vp p circuito abierto 10Vp p de...

Страница 25: ...a 2Mhz Nive 4V a 14 V 0 5Vp p continuamente variable Tiempo de elevaci n 20nS VCG GENERADOR DE VOLTAJE CONTROLADOR ENTRADA Voltaje de entrada 0 10V 1V causa un 100 1 cambio de frecuencia Impendencia 1...

Страница 26: ...compensaci n cd esta disponible en este jack 7 TTL CMOS jack TTL o CMOS onda cuadrada dependiendo en la posici n de el switch 13 de el nivel CMOS sale de este jack Esta salida es independiente del NI...

Страница 27: ...ntador cambia Cuando los 50KHz a 5MHz rangos son seleccionados el LED relampaguea 10 veces por segundo cada 0 1 segundos Cuando el 50 hasta 5K rango es seleccionado la LED relampaguea una vez cada seg...

Страница 28: ...e la onda de forma deseada seno cuadradazo triangulo oprimiendo uno de los switches de la FUNCION 3 Relaciones de fase de las formas de onda son mostradas en la figura 2 4 Seleccione la frecuencia de...

Страница 29: ...el control FINO no alcanzara su limite cuando se trate de finalizar el arreglo de la frecuencia 3 El control fino de frecuencia provee aproximadamente 5 desviado de frecuencia desde el GRUESO control...

Страница 30: ...ado es producida El Modelo 4011A provee por la variaci n de simetr a desde 15 hasta 85 1 Seleccione la onda de forma deseada sea SENO CUADRADA o TRIANGULAR 2 Enganche el CICLO DE SERVICIO switch 14 y...

Страница 31: ...o CW contra el reloj rotacion Por ejemplo si la frecuencia de comiezo es 500 KHz aplicando 10 V cambiara la frecuencia de salida a 5KHz 3 Para usar el generador de funciones como un generador de barri...

Страница 32: ...a y ajuste el NIVEL CMOS control como sea instruido en el manual 5 Cuando el generador de funciones es usado por estudiantes u otros usuarios sin experiencia el circuito mostrado en Fig 5 puede ser an...

Страница 33: ...mientos de servicio REEMPLAZO DE FUSIBLES 1 Localice el sostenedor de fusible en la linea de entrada 2 Quite el sostenedor del fusible y dubstituya el fusible por un fusible de valor egual SERVICIO DE...

Страница 34: ...t cnica sobre aplicaciones especiales de su instrumento Asesor a t cnica para escoger el mejor instrumento para una tarea espec fica Informaci n sobre accesorios opcionales para su instrumento Informa...

Страница 35: ...forma de queque orden de dinero o numero de carta de cr dito Para los precisos mas corriente visite www bkprecision com y oprime service repair Vuelva toda la mercanc a a B K Precision Corp con el en...

Страница 36: ...e quince 15 d as de la compra de este producto Exclusiones Esta garant a no se aplica en el evento de uso en error o abuso de este producto o el resultado de alteraciones desautorizado o reparaciones...

Страница 37: ...a estar en el lado seguro trate todos los dos cables ca equipo como chasis caliente al menos que este seguro que tiene un chasis aislado o un chasis de tierra 2 En instrumentos de prueba o culaquier e...

Страница 38: ...ers 4010A 4011A 4012A 4040A 4017A complies with the essential requirements of the following applicable European Directives Low Voltage Directive 73 23 EEC 19 02 73 amended by 93 68 EEC 22 07 93 Electr...

Страница 39: ...20 22820 Savi Ranch Parkway Yorba Linda CA 92887 4610 2005 B K Precision 480 844 9 001 Printed in U S A...

Отзывы: