7
Trigger Sensitivity:
AC:
1.0 div (internal)
≥
0.5 Vp-p (external)
30 Hz – 20 MHz
TV-V:
1.0 div (internal)
≥
0.5 Vp-p (external)
20 Hz – 20 kHz
TV-H:
1.0 div (internal)
≥
0.5 Vp-p (external)
3 kHz – 100 kHz
Maximum External Trigger Voltage:
300 V (dc +
ac peak).
DIGITAL STORAGE FACILITIES
Storage Word Size:
2048 x
8 bits/channel (2 k/channel with direct
sampling, 1 k/channel with equivalent time
sampling).
Vertical Resolution:
8 bit (1 in 256) approximately 25 steps/div.
Horizontal Resolution:
1 in 2048, approximately 200 samples/div.
Sampling Rate:
10 M samples/sec to 4 samples/sec, reduced in propor-
tion to time base. Direct sampling at time base settings
of 20 s/div and slower, equivalent time sampling at
time base settings of 10 µs/div and faster.
Time Base Expander:
For storage of slow time events, time base steps 10
ms/div and slower have selectable 1/1 or 1/100 rate.
1/100 rate expands time base from 1 sec/div to 50
sec/div in 1-2-5 sequence.
Equivalent Time Sampling Bandwidth:
20 MHz for repetitive waveforms.
Dot Joining:
Linear interpolation between samples.
Digital Display Modes:
Roll:
Stored data and display updated continually.
Refresh:
Stored data and display updated by triggered sweep.
Hold:
Freezes channel 1 and channel 2 data immediately.
Save CH2:
Freezes channel 2 data immediately.
Pretrigger Storage:
Available in single sweep mode, switchable to 0%,
25%, 50%, or 75%.
USB Host Interface
Save screen image to USB flash drive
OTHER SPECIFICATIONS
Cal/Probe Compensation Voltage:
0.5 V p-p ± 3% square wave, 1
kHz nominal.
CH 2 (Y) Output:
Output Voltage:
50 mV/div (nominal into 50 ohm load).
Output Impedance:
Approximately 50 ohms.
Frequency Response:
20 Hz to 20 MHz, - 3 dB.
Power Requirements:
100–130 VAC or 200–260 VAC, 5 0/60 Hz,
55 watts.
Dimensions (H
3
W
3
D):
5.2”
X
12.8”
X
15.7”
(132
X
324
X
398 mm).
Weight:
18.7 lbs (8.5 kg).
Environment:
Within Specified Accuracy:
+10° to +35° C, 10–80% relative
humidity.
Full Operation:
0° to +50° C, 10–80% relative humidity.
Storage: -
30° to +70° C, 10–90% relative humidity.
ACCESSORIES SUPPLIED:
Two Switchable X1/X10 Probes.
Instruction Manual.
AC Line Cord.