8560 and 8660 • Puncture and Biopsy Facilities

Fig. 4.

Mounting the puncture attachment UA1239 (shown without cover)

To alter the angle of the needle-guide holder:

1. Loosen the adjustable screw by turning it one quarter revolution counter-clockwise.

2. Select the desired angle (60

°

; 45

°

; 30

°

).

3. Retighten the adjustable screw.

To select the correct needle barrel:

1. Raise the three winged handle up out of its socket and rotate it until the desired barrel 

is next to the small black triangular marking on the top of the needle guide holder.

2. Push the three winged handle fully back down into its socket.

3. Slide in the desired needle (or catheter).

To detach the needle during interventional procedures:

1. Raise the three winged handle approx. 3mm and rotate counter-clockwise through 60

° 

until halted by the small metal nub. 

2. The needle (or catheter) can now be released from the needle guide attachment.

All parts of the puncture attachment can be autoclaved or disinfected by immersion in a 
suitable solution.

Содержание Type 8560

Страница 1: ...English BB0612 G August 2006 Types 8560 and 8660 Linear Array Transducer User Guide...

Страница 2: ...nd services As part of our customer satisfaction program we contact a sample of our customers a few months after they receive their orders If you receive an email message from us asking for your feedb...

Страница 3: ...fection 2 Starting Scanning 3 Connecting the Transducer 3 Changing Frequency 3 Using a Transducer Cover 3 Using the Transducer Control Button 3 Changing Orientation 4 Puncture and Biopsy Facilities 4...

Страница 4: ......

Страница 5: ...intraoperative use FDA WARNING for the United States of America Types 8560 and 8660 are not for fetal use Fig 1 Linear Array Transducer Types 8560 and 8660 Scanning Plane Fig 2 Scanning plane of Types...

Страница 6: ...ucer can withstand are listed in the Product Data sheet Sterile covers are available See the Product Data sheet for more details Caution Keep all plugs and sockets absolutely dry at all times Starting...

Страница 7: ...a use only licensed probe sheaths Note For Intraoperative Applications Apply sterile scanning gel to the tip of the transducer or fill the cover with 1 to 2ml of sterile water This improves the screen...

Страница 8: ...edles There is also a screw for adjusting the angle of penetration The needle guide will accept 0 9mm 20 Gauge 1 3mm 18 Gauge and 2 1mm 14 Gauge needles Fig 3 Puncture guide attachment UA1239 The brac...

Страница 9: ...s socket and rotate it until the desired barrel is next to the small black triangular marking on the top of the needle guide holder 2 Push the three winged handle fully back down into its socket 3 Sli...

Страница 10: ...n as a bright dot on the screen To remove the puncture line from the scan image refer to the applicable scanner user guide for instructions WARNING It is essential for the patient s safety that only t...

Страница 11: ...d from all needle guides and other channels and grooves See Transducer Care Cleaning Safety for cleaning instructions Disposal When the transducer is scrapped at the end of its life national rules for...

Страница 12: ......

Отзывы: