8823 User Guide
(BB1740-C)
9
1
Mount the attachment bracket on the transducer. Fit the indentations in the sides
of the bracket over the raised metal knobs on either side of the transducer handle.
2
Fix the attachment bracket in position by tightening the attachment lock screw
(item A in Figure 3). Tighten the screw up to its “locking point”, after which it
will not be possible to tighten the screw anymore.
3
The biopsy attachment should now be fixed solidly to the transducer’s handle.
4
Using the needle guide locking screw (item B in Figure 3), adjust the needle
guide according to the size of needle to be used.
The distance from the guide channel entrance of the puncture attachment to the
first dot on the scan image puncture line is 52 mm. The distance between the
dots is 10 mm, see Figure 8 for details.
Figure 4. 8823 with UA 1250 mounted.
All parts of the puncture attachment UA1250 can be autoclaved or disinfected
by immersion in a suitable solution. See
Care, Cleaning & Safety
that
accompanies this user guide.
UA1341
The puncture attachment kit comprises a reusable plastic bracket (UA1341) and
sterile, single-use needle guides (UA0013). Be careful not to discard the
reusable bracket together with the single-use needle guides. The bracket can be
disinfected by immersion in a suitable solution
.
To mount puncture attachment UA1341 ready for use:
1
Attach the bracket (UA1341) to the transducer.
2
Pull a transducer cover over the transducer and bracket.
WARNING
Sterile needle guides are disposable products intended for single use only.