WARNING
Ensure that the correct fuse type has been fitted in the mains
inlet fuse holder for the desired voltage selector switch value.
See Table 5 for details.
1. Remove the AC plug
from the power inlet
module
2. Use a small flat bladed
screwdriver to carefully
pull out the fuse holder
3. Fit the correct fuse in the
far position only
4. Push the holder back
into the inlet mode
Install and test the
xxsealer
unit using the following procedure:
1) Place the unit on the suitably selected working surface (as
specified above), ensuring that the ventilation holes on the
rear of the unit are not obstructed.
2) Connect the unit to the AC mains power outlet using the
mains power cord supplied or as specified in Section 4.
3) Switch the mains power on at supply outlet first, and then
switch the unit on using the power switch located at the rear of
the unit.
4) Check that the unit is stable and safe by performing a trial
sealing operation, as described in Section 7.2.
DANGER
If the unit has been stored in a cool environment, it must be
left to stand unplugged until it has acclimatised to the new
room temperature before being powered.