Operating instructions iS10 / iS20
Service menu
38019417000 en
79
9
Selection
Subselection
Display
Explanation
or
04 00
Battery pack switch-off time
15
Time in minutes after which the device automatic
switches off.
0
= Function not active
or
05 0
Weight change required for new recording
0
Function not active
Setting changes with
each keystroke.
1
Relative deviation from latest registered value
(input via step 39)
2
Relative deviation from latest registered value
(input via step 39)
and
Absolute deviation from zero cross-over
(input via step 40)
or
06 0
Reference weight filter
0
0.5 s
Setting changes with
each keystroke.
1
1.0 s
to
9
25 s
or
07 0
Auto recording
Requirement for this is: The scale is in equilibri-
um and within tolerance limits.
0
Function not active
1
Auto recording for tolerance control
or
08 0
Software update
0
Operating mode weighing operation
1
Operating mode software update
or
09 0
Load factory data
(EDP/Printer/General settings/Battery status)
0
Function not active
1
Factory data is loaded