Sleep&Go
User’s Manual
534-7AB-MU2 • REV. 1.07
User
’s Manual -
Cardior
espiratory polygraph - Sleep&Go
41
• Select the “TIME” field and set the starting time for the test.
Select “OK” when finished (E).
• Move to the “DURATION” field and enter the number of hours
and minutes for the test (F).
•
Select the option “CHANNELS CONF.”. The device displays all
channel configurations stored on the memory card. A channel
configuration establishes which channels are recorded during
the test, as well as their sampling rates. There may be up to 10
different channel configurations. A detailed explanation on how
to change these configurations is given below. Select which
configuration is more suitable for the type of test you want to
do (G).
• Select “SAVE” once all parameters are configured. You may
switch off now the Sleep&Go.
•
The device will start automatically the test at the designated
date and time. The screen will light for 50 seconds, displaying
the signals acquired by the sensors. The front panel LED will
blink during the whole test.