5. Safety & Maintenance
5.1. Electromagnetic Emissions
See the table below (Table 6) for specific information regarding the BITalino (r)evolution Board
electromagnetic emissions. This device is intended for use in the electromagnetic
environment(s) specified below. The user of this device must ensure that it is used within the
specified environment(s).
Table 6: Electromagnetic emissions. The device is suitable for use in all establishments, including domestic and
those directly connected to low-voltage energy public network supply which supplies buildings used for domestic
purposes.
Emission Tests
Compliance
Electromagnetic
Environment
Guidance
RF Emissions | CISPR 11
Group 1
Class B
This device must emit
electromagnetic energy
to perform its function.
Electronic equipment
near it, may be affected.
5.2. Safety Instructions
Please read the following safety instructions
before
using your
BITalino
system to prevent any
damages or problems with the user, test persons and/or
BITalino
devices. Violations of these
instructions can lead to inferior signal quality and/or damages to the
BITalino
system and user.
Warning & Precautions
Do not use this device in patients with implanted electronic devices
of any kind, including pacemakers, electronic infusions pumps,
stimulators, defibrillators or similar.
Do not apply electrodes over damaged skin.
Do not use the system in patients with allergies to silver
See all the documentation that came with the system for
instructions on safety and precaution. Also available at
https://bitalino.com/documentation#documention---tabs
Do not use this equipment in an environment with operating
magnetic imaging devices (MRI).
Do not use the device near the fire or in potentially explosive
atmosphere such as atmospheres with flammable gas.
The use of damaged accessories or not recommended by PLUX
may result in changes to the technical characteristics and features
of your BITalino (r)evolution Board. Use only the approved
accessories purchased from PLUX or a PLUX agent.