Directional Power Sensors

Operation Manual

Model 5009, 5010B, and 5014

©Copyright 2017

 by Bird Technologies Inc. 

Instruction Book Part Number 920-5010S Rev. E

Содержание 5009

Страница 1: ...Directional Power Sensors Operation Manual Model 5009 5010B and 5014 Copyright 2017 by Bird Technologies Inc Instruction Book Part Number 920 5010S Rev E ...

Страница 2: ...d Do not attempt to remove the RF transmission line while RF power is present WARNING Do Not Service Or Adjust Alone Under no circumstances should any person reach into an enclosure for the purpose of service or adjustment of equipment except in the presence of someone who is capable of rendering aid WARNING Safety Earth Ground An uninterruptible earth safety ground must be supplied from the main ...

Страница 3: ...jury CAUTION Caution notes call attention to a procedure which if not correctly performed could result in damage to the instrument WARNING Leaking RF energy is a potential health hazard DO NOT connect or disconnect equipment from the transmission line while RF power is being applied Severe burns electrical shock or death can occur WARNING Never attempt to connect or disconnect RF equipment from th...

Страница 4: ...ECCIÓN DE SEGURIDAD DEL INSTRUMENTO BENUTZUNG WIRD DAS GERÄT AUF ANDERE WEISE VERWENDET ALS VOM HERSTELLER BESCHRIEBEN KANN DIE GERÄTESICHERHEIT BEEINTRÄCHTIGT WERDEN UTILISATION TOUTE UTILISATION DE CET INSTRUMENT QUI N EST PAS EXPLICITEMENT PRÉVUE PAR LE FABRICANT PEUT ENDOMMAGER LE DISPOSITIF DE PROTECTION DE L INSTRUMENT IMPIEGO QUALORA QUESTO STRUMENTO VENISSE UTILIZZATO IN MODO DIVERSO DA CO...

Страница 5: ...GERÄTES GELTEN NUR FÜR GESCHULTES FACHPERSONAL ZUR VERMEIDUNG GEFÄHRLICHE ELEKTRISCHE SCHOCKS SIND WARTUNGSARBEITEN AUSSCHLIEßLICH VON QUALIFIZIERTEM SERVICEPERSONAL DURCHZUFÜHREN ENTRENTIEN L EMPLOI DES INSTRUCTIONS D ENTRETIEN DOIT ÊTRE RÉSERVÉ AU PERSONNEL FORMÉ AUX OPÉRATIONS D ENTRETIEN POUR PRÉVENIR UN CHOC ÉLECTRIQUE DANGEREUX NE PAS EFFECTUER D ENTRETIEN SI L ON N A PAS ÉTÉ QUALIFIÉ POUR C...

Страница 6: ... ÉLÉMENT H F CONSERVER L ÉLÉMENT DANS LA PRISE LORS DE L EMPLOI HF SPANNUNG KANN IN DER HF ELEMENT BUCHSE ANSTEHEN ELEMENT WÄHREND DES BETRIEBS EINGESTÖPSELT LASSEN PUEDE HABER VOLTAJE RF EN EL ENCHUFE DEL ELEMENTO RF MANTENGA EL ELEMENTO EN EL ENCHUFE DURANTE LA OPERACION IL PORTAELEMENTO RF PUÒ PRESENTARE VOLTAGGIO RF TENERE L ELEMENTO NELLA PRESA DURANTE IL FUNZIONAMENTO ...

Страница 7: ...ents to our Solon Ohio factory This manual may be periodically updated When inquiring about updates to this manual refer to the part number and revision on the title page Chapter Layout Introduction Describes the features of the Directional Power Sensor and Element Types Installation Describes how to connection and install the Directional Power Sensor into the system that is being monitored Specif...

Страница 8: ...ut vi Chapter 1 Introduction 1 Description 1 Element Types 1 43 Type Elements 1 APM DPM Elements 1 Element Orientation 2 Element Contact Alignment 2 Chapter 2 Installation 4 Unpacking and Inspection 4 Connecting the Directional Power Sensor DPS 5 Chapter 3 Specifications 8 5009 Specifications 8 5010B 5010T Specifications 10 5014 Specifications 12 Customer Service 14 Limited Warranty 15 ...

Страница 9: ... to measure peak power These elements can measure the peak power of a system with an accuracy of 8 of full scale as long as the signal meets the following requirements At least 15 pulses per second PPS Minimum pulse width of 15 µs 800 ns if frequency is greater than 100 MHz Minimum Duty Cycle of 0 01 In addition 43 type elements can be used to measure average power in signals with a peak to averag...

Страница 10: ... the contact spring changes position erratic power readings may be experienced The position of the contact spring may be adjusted with a small screwdriver to reestablish contact Perform the following steps to adjust the contact spring 1 Using a small flat head screwdriver place the flat side of the screwdriver behind the contact bar as indicated and bend the contact bar so that the contact rests i...

Страница 11: ...e element socket bore so that the profile of the element contact is visible when viewing the element socket from the top of the socket bore Note If the contact is accidentally moved too far the element will not slide into the socket Move the contact back into the recessed area and repeat the process ...

Страница 12: ...ntainer is damaged do not unpack the unit Immediately notify the shipping carrier and Bird Technologies If the shipping container is not damaged unpack the unit Save shipping materials for repackaging 2 Inspect unit for visual signs of damage Note If there is damage immediately notify the shipping carrier and Bird Technologies ...

Страница 13: ...DPS Note Only use DPM elements in the 5009 DPS WARNING Never attempt to connect or disconnect RF equipment from the transmission line while RF power is being applied Leaking RF energy is a potential health hazard WARNING RF voltage may be present in RF element socket Keep element in socket during operation ...

Страница 14: ...e Bird Digital Power Meter using the sensor cable provided Figure 2 5010 Cable Connection c To connect a 5014 DPS Note There are currently three connection options for the 5014 DPS Connect the DPS to a USB port on a PC using the sensor cable pro vided Use VPM3 software to communicate with the DPS Connect the DPS to the Sensor USB port on a Bird Digital Power Meter using the sensor cable provided C...

Страница 15: ... 2 Note The arrow on the forward element should point towards the load Note The arrow on the reflected element should point towards the source Note Both elements must be either APM DPM or 43 types do not mix elements 3 Refer to DPM VPM3 CPM manuals for setup parameters for the desired measurement Note Use the forward element s power rating when setting up the device ...

Страница 16: ...turn Loss Rho VSWR 0 to 20 dB 0 1 to 1 1 22 to 99 99 Uncertainty Twice the Avg Power Uncertainty Calculated from forward and reflected uncertainty Settling Time Max 2 5 seconds Impedance Nominal 50 ohms Insertion Loss Max 0 05 dB up to 1 GHz Input VSWR Max 1 05 1 up to 1 GHz Directivity Typical 1 30 dB RF Connectors QC Type N F normally supplied Interface Connector RJ 25 Power Supply From host ins...

Страница 17: ...emp Storage 40 to 75 C 40 to 167 F Humidity Max 95 non condensing Altitude Max 3 000 m 10 000 ft Dimensions Nominal 5 0 x 2 4 x 2 0 130 x 60 x 50 mm Weight Nominal 0 9 lb 0 4 kg 1 Exact value depends on element selected ...

Страница 18: ...e 5010T only 230 ms 43 Elements Forward or Reflected Direction Power Measurement Range 0 1 W to 10 kW Power Measurement Accuracy 5 of full scale average power 95 c l Peak Power Measurement 43 Elements only Forward direction only Pulse Width Min 2 25 MHz 25 100 MHz 100 MHz 15 µs 1 5 µs 800 ns Rep Rate Min 15 PPS Duty Cycle Min 1 x 10 4 Accuracy 2 8 of full scale peak envelope power 95 c l Match Mea...

Страница 19: ...In accordance with MIL PRF 28800F Class 3 CE CE compliant Refer to DOC for specific standards Recommended Calibration Interval 1 year Temp Operating 10 to 50 C 14 to 122 F Temp Storage 40 to 75 C 40 to 167 F Humidity Max 95 non condensing Altitude Max 3 000 m 10 000 ft Dimensions Nominal 5 0 x 2 4 x 2 0 130 x 60 x 50 mm Weight Nominal 0 9 lb 0 4 kg 1 Exact value depends on element selected 2 Above...

Страница 20: ...rection Power Measurement Range 0 1 W to 10 kW Power Measurement Accuracy 5 of full scale average power 95 c l Peak Power Measurement 43 Elements only Forward direction only Pulse Width Min 2 25 MHz 25 100 MHz 100 MHz 15 µs 1 5 µs 800 ns Rep Rate Min 15 pps Duty Cycle Min 1 x 10 4 Accuracy 2 8 of full scale peak envelope power 95 c l Match Measurement Match Range Return Loss Rho VSWR 0 to 20 dB 0 ...

Страница 21: ...e with MIL PRF 28800F Class 3 CE CE compliant Refer to DOC for specific standards Recommended Calibration Interval 1 year Temp Operating 10 to 50 C 14 to 122 F Temp Storage 40 to 75 C 40 to 167 F Humidity Max 95 non condensing Altitude Max 3 000 m 10 000 ft Dimensions Nominal 5 0 x 2 4 x 2 0 130 x 60 x 50 mm Weight Nominal 0 9 lb 0 4 kg 1 Exact value depends on element selected 2 Above 35 C or bel...

Страница 22: ...o be returned for any reason request an Return Material Authorization RMA through the Bird Technologies website All instruments returned must be shipped prepaid and to the attention of the RMA number Bird Service Center 30303 Aurora Road Cleveland Solon Ohio 44139 2794 Fax 440 248 5426 E mail bsc birdrf com For the location of the Sales Office nearest you visit our Web site at http www birdrf com ...

Страница 23: ...ry such claim for breach of these warranties shall be deemed to be waived by Buyer unless made in writing within ten 10 days from the date of discovery of the defect The above warranties shall not extend to any products or parts thereof which have been subjected to any misuse or neglect damaged by accident rendered defective by reason of improper installation or by the performance of repairs or al...