58
Duty Cycle —
Setting is a user selectable value between 100 and 0.
auto enabled
Bandwidth (Video Filter) —
See Bandwidth (Video Filter) under Time
Domain Measurement.
Run Mode —
Normal, RF Meter takes one measurement and stops
Clear/Restart, Once one measurement is completed and displayed
another measurement is started automatically.
Confidence Percentage —
Setting is a user selectable value of: 80, 90, 95,
99, 99.9, or 99.99%.
Sample Time —
Setting is a user selectable value of 0.0261 (26.1 ms) to
6.5535 seconds.
Sample Count —
Setting is a user selectable value of 1144900 to 288354000
samples.
Dynamic Range —
Selects the x-axis chart scale, setting is a user selectable
value 0 - 43 dB.
Zeroing Sensor —
Over time, the sensor's "zero value" (reading with no
applied RF power) can drift, making all readings inaccurate by this value.
For example, if the zero value is 0.02W, measuring a 50 W signal will give a
reading of 49.98 W, a 0.04% error. Measuring a 1 W signal will give a reading of
0.98 W, a 2% error. If the drift would be a significant error, zero the sensor.
Confidence Percentage
Sample Time (seconds)
Sample Count
80%
0.0261
1148400
90%
0.0339
1491600
95%
0.0421
1852400
99%
0.0604
2657600
99.9%
0.6038
26571600
99.99%
6.03850
265694000